Description
ABS PH-ABT-HC-RFC12A-CAD Pharmacy Refrigerator Freezer
Our ABS PH-ABT-HC-RFC12A-CAD Pharmacy Refrigerator Freezer with the patented “Controlled Auto Defrost” Technology provides the convenience of an auto-defrost design while still ensuring that vaccines remain at consistent and stable temperatures below 5°F (-15°C)
The acceptable temperature ranges for vaccine storage are between 36° Fahrenheit and 46° Fahrenheit (or between 2° Celsius and 8° Celsius) in the refrigerator and between -58° Fahrenheit and +5° Fahrenheit (or between -50° Celsius and -15° Celsius) in the freezer.
Pharmacy Refrigerator Freezer Features:
- Refrigerator/freezer freestanding thermometer with 3 year certificate of calibration, with min/max memory, °f/°c switchable
- Audible and visual high and low temperature alarms, refrigerator and freezer
- Remote alarm contacts, refrigerator and freezer
- Six adjustable shelves with guard rail on back to prevent product from being pushed against back wall, refrigerator
- Two adjustable shelves with guard rail on back to prevent product from being pushed against back wall, freezer
- Bottom footed shelf to allow air circulation, refrigerator and freezer
- Vaccine storage power cord warning label
- Forced draft circulation refrigerator
- Keyed door locks
- Magnetic door gasket for positive seal
- Integrated door handles
- Leveling legs
- Stacking kit
- Two probe access ports (3/8″); one, refrigerator/one, freezer
- White powder coated exterior
- Hfc-free refrigerant (r600a), foam insulation and packaging
- Pharmacy refrigerator/freezer toolkit and temperature logs
- Two year parts and labor warranty, plus an additional three year compressor parts warranty
- Pyxis®, omnicell ®and acudose rx® compatible
Pharmacy Refrigerator Freezer Specifications:
- Exterior dimensions: 23 ¾”w x 26 ½” d x 80 ¼”h
- Controlled auto defrost freezer, patented technology
- Cycle defrost refrigerator
- 115v, 60 hz, 2.3 amps refrigerator, 1 amp freezer, 1/3 hp refrigerator, 1/8 hp freezer
- Ul/c-ul listed
- Shipping weight: 289 lbs.
- In stock, with next day shipping
- *please note this unit requires 2 outlets.
Pharmacy Refrigerator Freezer Temp Controller:
- Microprocessor temperature controller, refrigerator and freezer
- Digital temperature display with sample simulated probe, refrigerator and freezer
- Adjustable operating temperature range:
- Refrigerator: 36°f – 46°f (2°c – 8°c)
- Freezer: 5°f – -13°f (-15°c – -25°c)
- Defrost:
- Auto defrost cycle every 6 hours: (at set point -11°f /-24°c)
- Max air temp in empty freezer: 5°f (-15°c)
- Max air temp in loaded freezer: 3°f (-16°c)
- Max sample temp in empty freezer: 0°f (-18°c)
- Max sample temp in loaded freezer: -4°f (-20°c)
American Biotech Supply Refrigeration Solutions to Meet Tomorrow’s Standards. ABS Offers you creative product designs and an open mind for customer specific applications. View our wide selection of Premier Laboratory Refrigerators and Freezers.
We have a vast selection of laboratory refrigeration and freezer models to choose from, ranging down to Ultra-Low -86°C cascade freezers and -196°C cryogenic freezers.
Product FAQ
What is "Controlled Auto Defrost" technology?
This patented technology provides the convenience of an automatic defrost system while specifically managing freezer temperatures to ensure they remain consistent, stable, and well below the 5°F (-15°C) threshold, which is critical for vaccine and sensitive sample integrity.
What temperature security features are included?
The unit features microprocessor controllers for both refrigerator and freezer sections with digital displays. It includes an alarm package with audible and visual high/low-temperature alerts, a power failure alarm, sensor error monitoring, and remote alarm contacts. Two probes (one in air, one in a sample-simulated liquid bottle) are included for each section.
Is this unit compatible with pharmacy management systems?
Yes, it is compatible with major pharmacy tracking and documentation systems like Pyxis®, Omnicell®, and AcuDose-Rx®.
What is the warranty coverage?
It includes a standard two-year parts and labor warranty, plus an additional three-year compressor parts warranty for reliable, long-term operation in medical environments.
Shipping FAQ
How is this refrigerator/freezer combination shipped?
Due to the unit’s specialized electronics and internal cooling system, it is shipped via LTL (Less-Than-Truckload) freight, securely palletized and crated to minimize movement and ensure it arrives in optimal condition. Please note that this unit requires two separate power outlets for operation.
Do I need liftgate service?
Yes, liftgate service is mandatory unless your facility has a loading dock capable of receiving a large-scale transport vehicle, as this unit must be carefully lowered to ground level.
What should I inspect upon arrival?
Before signing the delivery receipt, you must perform a comprehensive inspection of the unit, specifically focusing on the exterior panels, door seals, and the refrigeration housing for any signs of transit-related damage.
What are the post-delivery setup steps?
To prevent damage to the compressor and ensure long-term cooling stability, you must allow the unit to sit upright in its final installation location for at least 24 hours to let all fluids settle before powering it on.