Description
ABS PH-ABT-NSF-UCBI-0404G-ADA-LH Vaccine Refrigerator
These premier ABS PH-ABT-NSF-UCBI-0404G-ADA-LH Vaccine Refrigerator are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.
The glass door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.
*-one or more of these certifications may apply to this unit.
Vaccine Refrigerator Features:
- Description– single glass door pharmacy/vaccine undercounter refrigerator built-in ADA compliant
- Operational environment– indoor use only, +18°c to +26°c (+65°f to +78°f), <70% rh
- Storage capacity– 4.6 cu. Ft. Gross volume
- Door– one swing glass door, self-closing, left hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves– three shelves (two adjustable/one fixed) with guard rail on back
- Mounting– low profile roller wheels and leveling legs
- Interior lighting– na
- Airflow management– forced air technology, patent pending
- External probe access– rear wall port (1/2″) dia.
- Insulation– cabinet is foamed-in-place with epa compliant high density urethane foam
- Exterior materials– white powder coated steel
- Access control– pyxis®, omnicell® and acudose rx® compatible
- General warranty– two (2) years parts and labor warranty, excluding display probe calibration
- Compressor warranty– five (5) years compressor warranty
- Exterior width– 23 7/8″
- Exterior depth– 26″
- Exterior height– 31 15/16″
- Product weight– 100 lbs.
- Shipping weight– 140 lbs.
- Rated amperage– 1.74 amps
- Power plug/power cord– nema 5-15 plug, 8 to 10 ft typical, conforms to ul471 requirements, vaccine storage power cord warning label
- Facility electrical requirement– 110-120v ac: 15 a (minimum)
- Agency listing and certification– certified in accordance with the nsf/ansi 456 standard for vaccine storage. Ul, c-ul, etl, c-etl listed (either single or dual agency listings) and certified to ul471 standard, hydrocarbon refrigerant safety.
- Included accessories– temperature monitor device (tmd) complies with the current cdc guidelines, with 3 years certification of calibration, “buffered” probe in the product simulated solution, min/max memory. Field installable, and visual & audible temp alarm
Pharmacy refrigerator/freezer toolkit and temperature logs
Vaccine Performance Refrigeration system :
- Compressor– hermetic, high performance
- Refrigerant– epa snap compliant, r600a, isobutane
- Condenser– hybrid fin and tube with low noise fan
- Evaporator– plate wall
- Defrost– cycle optimized, zero energy
Performance:
- Uniformity¹ (cabinet air)– +/- 0.8°c
- Stability² (cabinet air)– +/- 1.2°c
- Maximum temperature variation (cabinet air)– +/- 1.4°c
- Temperature rise after an after 8 sec door openings– temperature did not exceed 6.4°c at any probe for all required nsf/ansi 456 testing protocolsᶟ
- Recovery after 3 min door opening– all probes recover to under 8°c within 4.8 min.
- Energy consumption– 1.15 kwh/day⁴
- Average heat rejection– 1.57 kwh/day (224 btu/h)⁴
- Noise pressure level (dba)– 43 or less installed
- Pull down time to nominal operating temp– 35 min
Performance data acquired at 22°c ambient, using nsf/ansi 456 compliant validation ballast probes, empty chamber, during stabilized steady state operation and a daq sampling rate of one measurement every 10 seconds
Uniformity is defined as the maximum variance in temperature across all probes at any point in time over the testing period
Stability is defined as the maximum variance in temperature experienced by any single probe over the testing period
Temperature performance for all loaded and unloaded door opening protocols, all alarm, controller and probe requirements as defined in the nsf/ansi 456 standard for vaccine storage
Data per energy star test results or equivalent testing and calculation. Heat rejection based on daily averages, not continuous operation. Performance exceeds energy star requirements.
Vaccine Temp Controller:
- Controller technology– parametric, microprocessor, led display with 0.1°c resolution
- Temperature setpoint range– 1°c to 10°c (setpoint must remain unaltered from the factory setting to remain compliant with nsf/ansi 456 standard for vaccine storage requirements)
- Display probe– calibrated, stainless steel
- External alarm connection– state switching remote alarm contacts
- Alarms– visual and audible indicators, High / low temperature, compliant with alarm requirements defined in the nsf/ansi 456 standard for vaccine storage
- Simulator ballast– glass bead thermal media
American Biotech Supply Refrigeration Solutions to Meet Tomorrow’s Standards. ABS Offers you creative product designs and an open mind for customer specific applications. View our wide selection of Premier Laboratory Refrigerators and Freezers.
We have a vast selection of laboratory refrigeration and freezer models to choose from, ranging down to Ultra-Low -86°C cascade freezers and -196°C cryogenic freezers.
Product FAQ
What makes this model ADA-compliant, left-hinged, and glass-doored?
This specialized unit features a lower overall height profile to fit beneath ADA-compliant counters, a factory-installed left-hinged swing configuration to accommodate strict layout constraints, and a high-visibility glass door for quick inventory audits without thermal disruption.
What does NSF/ANSI 456 certification mean for this vaccine refrigerator?
NSF/ANSI 456 is the rigorous standard for vaccine storage units. This certification guarantees that the refrigerator and freezer maintain strict temperature performance, uniformity, and stability required by the CDC for safe vaccine storage.
Can this unit be built into cabinetry or counters?
Yes. The front-breathing, undercounter built-in design allows the unit to fit seamlessly beneath standard or lower counters without requiring the side or rear ventilation clearances demanded by standard freestanding models.
What temperature monitoring features are included?
The unit features independent microprocessor controllers, digital temperature displays, and a robust alarm suite with audible and visual high/low-temperature alerts, open-door warnings, and remote alarm contacts. It includes NIST-traceable calibration with sample-simulated buffered probes.
What is the warranty coverage?
It is backed by a standard parts and labor warranty, along with extended compressor coverage to ensure long-term clinical reliability.
Shipping FAQ
How is this undercounter medical refrigerator shipped?
To prevent transit damage to the glass door, sensitive internal cooling components, and precise digital electronics, this unit is shipped via LTL (Less-Than-Truckload) freight, securely palletized.
Do I need liftgate service?
Yes, liftgate service is required unless your facility has a loading dock capable of receiving a major transport vehicle, ensuring the unit is lowered safely to street level.
What should I inspect upon arrival?
Before signing the delivery slip, thoroughly inspect the glass door, exterior packaging, handles, and seals for any indication of transit-related impact.
What are the post-delivery setup steps?
Allow the refrigerator to sit completely upright in its final installation space for at least 24 hours prior to powering it on, letting all internal cooling fluids settle properly.