Description
Summit Medical Healthcare Refrigerator Description:
EQTemp introduces a new series of affordable, specialty medical refrigeration solutions for healthcare facilities seeking specialized cold storage.
Model ACR1011W is an upright 9.88 cu.ft. all-refrigerator. It features a clean white exterior and a blue-accented control panel. The solid door is equipped with a self-closing mechanism to help protect stored items from being overexposed to room temperature. A factory-installed lock offers added security, and the hospital-grade cord with a ‘green dot’ plug ensures a secure connection for high-traffic areas.
The Summit ACR1011W Medical Healthcare Refrigerator includes an external readout of the interior temperature and the high/low temperature. The buffered temperature probe is encased in a bottle filled with glass beads for improved accuracy. The digital temperature controls allow you to set the interior between 2 and 8ºC. An audible alarm sounds if the temperature goes out of range. We also include a pre-installed 3/8″ access port for users to add their own monitoring equipment if preferred.
The interior includes one wire shelf in the center to hold the temperature probe, as well as four additional adjustable shelves to accommodate different sizes. Internal LED lighting provides an efficient and convenient view. Automatic defrost reduces user maintenance, and an internal fan ensures robust temperature recovery after door openings.
This unit is also available certified to the NSF/ANSI 456 Vaccine Storage Standard with the upgrade of a NIST calibrated temperature readout (see model ACR1011WNSF456). Additional sizes are available, as well as more advanced cold storage equipment purpose-built for laboratory and life science storage. Browse the complete product line to explore our extensive selection.
Note: EQTemp’s ACR Series refrigerators are intended for various healthcare storage applications. Storage of high-value vaccines, life science materials, and medical/clinical biologicals requires a product offering tighter temperature control and/or alarms, monitoring, and backup.
Summit Medical Healthcare Refrigerator Features:
- EQTemp Healthcare Refrigerator– All-refrigerator designed to operate between 2 and 8ºC for a variety of storage application in medical offices and other healthcare facilities
- Temperature Range: 2º to 8ºC– Preset to 5ºC as recommended for many storage applications
- Large capacity– Full 9.88 cu.ft. capacity inside conveniently slim footprint
- External temperature readout– Includes continuous display of the current interior and high/low temperature, and can be switched from Celsius to Fahrenheit at the touch of a button
- High/low temperature alarm– Audible alarm sounds if the interior temperature rises or falls out of set range
- Buffered temperature probe– Thermometer probe is encased in a glass bead-filled bottle and stored in a fixed box on the center shelf. This is the ideal way to produce temperature readings closest to stored contents, as recommended by the CDC and VFC’s guidelines
- Access port– 3/8″ probe hole with cover is pre-installed in unit’s side to allow easy installation of user-provided monitoring equipment
- Factory installed lock– Keyed lock for a secure interior (2 keys included)
- Automatic defrost– Reduced user maintenance with automatic defrost
- Internal fan– Fan-forced cooling ensures the unit is evenly cooled from top to bottom
- Hospital grade cord with ‘green dot’ plug– Secure three-pronged cord for added unit safety in high traffic areas
- Fixed center shelf– Center shelf is fixed in place in order to provide the optimum position for the TMD to provide the most accurate temperature readings
- Flat door liner– No door storage to allow for the interior to accommodate deeper trays
- LED lighting– Attractively illuminates the interior in a gentle white hue for easy viewing
- Foamed-in-place cabinet– Quality design ensures efficiently insulated interior
- Temperature memory function– Unit will retain its last settings when power is restored in the event of power loss
- Self-closing door– Door will automatically close when left ajar, preventing items from accidentally being left exposed to ambient temperatures
- Pocket handle– Handle is recessed into the side of the door to ensure no added depth
- Open door alarm– Audible alarm sounds if the door is left open for more than one minute
- Digital thermostat– Digital microprocessor for easy and precise temperature management, with all controls externally located for added convenience
- Magnetic door gasket– Quality door gaskets ensure a tight seal for energy savings and improved performance
- Fully finished white cabinet– Allows the unit to be used freestanding
- CARB Compliant– Eco-friendly hydrocarbon-based refrigerant for a greener footprint that meets current CARB regulations
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
Summit Medical Healthcare Refrigerator Specifications:
- Height of Cabinet– 53.13″
- Width– 23.38″
- Width with Door Open– 23.88″
- Depth– 27.75″
- Depth with door at 90°– 45.25″
- Capacity– 9.88 cu.ft.
- Defrost Type– Automatic
- Door– White
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Amps– 2.0
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 116.8 lbs.
- Shipping Weight– 191.0 lbs.
- Parts & Labor Warranty– 1 Year
- Compressor Warranty– 5 Years
Summit Medical Healthcare Refrigerator Features:
- Door Swing– RHD
- Reversible– Factory Reversible
- Shelf Type– Wire
- Shelf Qty– 4
- Adjustable Shelves– Yes
- Thermostat Type– Digital
- Fan Type– Interior
- Refrigerant Type– R600a
- Refrigerant Amount– 2.0oz.
- High Side PSI– 1.45
- Low Side PSI– 0.85
- Level Legs Qty– 2
- Interior Light– Yes
- Temperature Range– 2º to 8ºC
- Sound Level– 37.7 dB
Summit Medical Healthcare Refrigerator Dimensions:
- Interior Height– 44.63″
- Interior Width– 20.0″
- Interior Depth– 21.5″
- Compressor Step Height– 9.0″
- Compressor Step Width– 20.0″
- Compressor Step Depth– 8.0″
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the ACR1011W and what is it used for?
The ACR1011W is a healthcare-grade upright refrigeration unit (~9.9 cu ft / ~280 L) designed for clinical, pharmacy or laboratory settings. It maintains a controlled interior temperature in the +2 °C to +8 °C range for safe storage of vaccines, reagents or medical supplies.
What are the key technical specifications?
• Approximate capacity: ~9.9 cu ft (~280 L)
• Temperature range: +2 °C to +8 °C (preset ~+5 °C)
• Cabinet width: ~24″ (~61 cm) (24″ wide upright model)
• Electrical: 115 V AC / 60 Hz (check local compatibility)
• Fully finished white cabinet, self-closing solid door, hospital-grade “green-dot” plug
What safety and monitoring features are included?
Includes external digital read-out (current + high/low interior temperature), audible alarm for out-of-range or door-open, buffered temperature probe in a glass-bead bottle for accurate simulation, factory-installed keyed lock, and 3⁄8″ access port for external monitoring.
Which environments is this unit best suited for?
Ideal for full-height upright cold-storage in hospitals, vaccine clinics, pharmacies or lab environments needing certified +2 °C-+8 °C storage. Not intended for general food or break-room use.
What warranty coverage is provided?
Manufacturer warranty: 1 year parts & labor; compressor typically covered for 5 years.
Shipping FAQ
How long does shipping take after placing an order?
Typical preparation and dispatch time is **1-3 business days** when in stock; larger medical cabinets may require additional freight scheduling.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping is possible. For deliveries to Melbourne, Australia, consider import duties/customs, freight access and site clearance (especially for full-height units), and verify local plug/voltage compatibility (unit is rated for 115 V AC/60 Hz).
How is the unit packaged for transit and what site requirements should be met?
The unit ships boxed/palletized with protective packaging. Approx dimensions: ~135 cm H × ~61 cm W × ~72 cm D (estimate). Ensure delivery site has level surface, sufficient doorway/entry clearance including depth + door-swing, and safe handling equipment.
What should I do if my shipment arrives damaged or is delayed?
If the unit arrives with visible damage or delivery is significantly delayed: document the condition (photos recommended), retain packaging and delivery receipt, then contact the supplier with your order/tracking details to initiate a claim or replacement.
Application FAQ
How do I install and set up the ACR1011W?
1. Place the unit on a firm, level surface (or approved built-in space if permitted) with proper clearance for ventilation at the rear/top.
2. Connect to a dedicated grounded outlet matched to the unit’s electrical specification.
3. Set the thermostat to the target temperature (~+5 °C within the +2 °C to +8 °C range), allow the interior to stabilize (typically a few hours) before loading materials.
4. Verify that the temperature display, high/low alarm, door-open alarm, and monitoring access port are functioning correctly before normal operation.
What best practices should I follow when storing items?
• Use the provided shelving without blocking airflow or vents.
• Avoid over-packing the interior; allow space for air circulation.
• Group frequently accessed items at the front to reduce door-open time.
• Label contents clearly for traceability.
• Inspect door gasket/seal and lock regularly and keep a log of temperature and alarm events if required by your facility.
What should I do if a temperature excursion or power interruption occurs?
If the interior temperature moves outside the acceptable +2 °C to +8 °C range or power is lost:
• Record the event (time, temperature, items impacted).
• Move critical contents to backup cold storage if available.
• Identify root cause (e.g., door left open, power outage, sensor alarm).
• Once resolved, reset any alerts/controls and monitor until stable before resuming standard operation.
What maintenance and verification routines are recommended?
Recommended routines include: monthly inspection of door gaskets and hinges; quarterly cleaning of condenser and ventilation paths; periodic testing of alarms and monitoring probe functionality; maintain calibration/service logs as required for auditing and compliance.