Description
Summit Vaccine Healthcare Refrigerator Description:
EQTemp by Accucold introduces a new series of NSF/ANSI 456 certified refrigeration designed to store a variety of materials in healthcare facilities.
Summit ACR1602GNSF456LHD Vaccine Healthcare Refrigerator is an upright 15.53 cu.ft. all-refrigerator. It features a clean white exterior and a blue-accented control panel. The glass door is equipped with a self-closing mechanism to help protect stored items from being overexposed to room temperature. A factory-installed lock offers added security, and the hospital-grade cord with a ‘green dot’ plug ensures a secure connection for high-traffic areas. This unit features a front-breathing design for safe built-in installation under counters. The fully finished cabinet also allows freestanding use.
The Summit ACR1602GNSF456LHD Vaccine Healthcare Refrigerator includes an external readout of the interior temperature, as well as the high/low temperature as required for data records. The buffered temperature probe is encased in a bottle filled with glass beads for improved accuracy and calibrated to NIST standards in our in-house ISO/IEC 17025:2017 certified laboratory. The digital temperature controls allow you to set the interior between 2 and 8ºC. An audible alarm sounds if the temperature goes out of range. We also include a pre-installed 3/8″ access port for users to add their own monitoring equipment if preferred.
The interior includes one wire shelf in the center to hold the temperature probe, as well as additional adjustable shelves to accommodate different sizes. Internal LED lighting provides an efficient and convenient view. Automatic defrost reduces user maintenance and an internal fan ensures powerful temperature recovery after door openings. This unit ships with a left-hand door swing (for a right-hand version, see model ACR1602GNSF456).
EQTemp’s series of ACR refrigeration is certified to the NSF/ANSI 456 Vaccine Storage Standard. All models feature hydrocarbon refrigerant for a greener footprint. Additional sizes are available, as well as more advanced cold storage equipment purpose-built for laboratory and life science storage. Browse the full product line to explore our extensive selection.
Note: EQTemp’s ACR Series refrigerators are intended for a wide variety of healthcare storage applications. Storage of high-value vaccines, life science materials, and medical/clinical biologicals require a product offering tighter temperature control and/or alarms, monitoring, and backup.
Summit Vaccine Healthcare Refrigerator Features:
- EQTemp Healthcare Refrigerator– All-refrigerator designed to operate between 2 and 8ºC for a variety of storage application in medical offices and other healthcare facilities
- Certified to NSF/ANSI 456 Vaccine Storage Standard– Tested to comply with the latest standards
- Temperature Range: 2º to 8ºC– Preset to 5ºC as recommended for many storage applications
- Certificate of calibration included– The temperature monitoring device (TMD) is calibrated to NIST standards in our ISO/IEC 17025:2017 certified calibration laboratory and ships with a 2-year certificate of calibration for your records
- Large capacity– Full 15.53 cu.ft. capacity inside conveniently slim footprint
- External temperature readout– Includes continuous display of the current interior and high/low temperature, and can be switched from Celsius to Fahrenheit at the touch of a button
- High/low temperature alarm– Audible alarm sounds if the interior temperature rises or falls out of set range
- Buffered temperature probe– Thermometer probe is encased in a glass bead-filled bottle and stored in a fixed box on the center shelf. This is the ideal way to produce temperature readings closest to stored contents, as recommended by the CDC and VFC’s guidelines
- Access port– 3/8″ probe hole with cover is pre-installed in unit’s side to allow easy installation of user-provided monitoring equipment
- Factory installed lock– Keyed lock for a secure interior (2 keys included)
- Automatic defrost– Reduced user maintenance with automatic defrost
- Internal fan– Fan-forced cooling ensures the unit is evenly cooled from top to bottom
- Hospital grade cord with ‘green dot’ plug– Secure three-pronged cord for added unit safety in high traffic areas
- Fixed center shelf– Center shelf is fixed in place in order to provide the optimum position for the TMD to provide the most accurate temperature readings
- Flat door liner– No door storage to allow for the interior to accommodate deeper trays
- LED lighting– Attractively illuminates the interior in a gentle white hue for easy viewing
- Foamed-in-place cabinet– Quality design ensures efficiently insulated interior
- Temperature memory function– Unit will retain its last settings when power is restored in the event of power loss
- Self-closing door– Door will automatically close when left ajar, preventing items from accidentally being left exposed to ambient temperatures
- Pocket handle– Handle is recessed into the side of the door to ensure no added depth
- Open door alarm– Audible alarm sounds if the door is left open for more than one minute
- Digital thermostat– Digital microprocessor for easy and precise temperature management, with all controls externally located for added convenience
- Magnetic door gasket– Quality door gaskets ensure a tight seal for energy savings and improved performance
- Fully finished white cabinet– Allows the unit to be used freestanding
- Left hand door swing– For this unit with a standard right hand door swing, see model ACR1602GNSF456
- CARB Compliant– Eco-friendly hydrocarbon-based refrigerant for a greener footprint that meets current CARB regulations
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
Summit Vaccine Healthcare Refrigerator Specifications:
- Height of Cabinet– 76.75″
- Width– 23.38″
- Width with Door Open– 23.63″
- Depth– 28.0″
- Depth with door at 90°– 50.0″
- Capacity– 15.53 cu.ft.
- Defrost Type– Automatic
- Door– Glass
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Sanitation– ETL-S to NSF 456
- Amps– 2.0
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 176.4 lbs.
- Shipping Weight– 259.5 lbs.
- Parts & Labor Warranty– 1 Year
- Compressor Warranty– 5 Years
Summit Vaccine Healthcare Refrigerator Features
- Door Swing– LHD
- Reversible– Factory Reversible
- Shelf Type– Wire
- Shelf Qty– 6
- Adjustable Shelves– Yes
- Thermostat Type– Digital
- Fan Type– Interior
- Refrigerant Type– R600a
- Refrigerant Amount– 2.0oz.
- High Side PSI– 1.45
- Low Side PSI– 0.85
- Level Legs Qty– 2
- Temperature Range– 2º to 8ºC
Summit Vaccine Healthcare Refrigerator Dimensions:
- Interior Height– 68.25″
- Interior Width– 20.0″
- Interior Depth– 21.5″
- Compressor Step Height– 9.0″
- Compressor Step Width– 20.0″
- Compressor Step Depth– 8.0″
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the ACR1602GNSF456LHD and what is it designed for?
The ACR1602GNSF456LHD is a full-size upright healthcare-grade refrigerator with approximately 15.53 cu ft (≈ 440 L) of storage, a 24 inch (≈ 61 cm) width and a height of about 196 cm. It is certified for vaccine and clinical storage under stringent conditions of +2 °C to +8 °C.
What temperature range and control features does it support?
It maintains a temperature range of +2 °C to +8 °C, uses a digital microprocessor controller with external display including high/low logging, and includes a buffered temperature probe for accurate stored-product monitoring.
What alarm, monitoring and security features are included?
Features include audible and visual high/low temperature alarms, door-ajar alert, power-failure warning, remote alarm contact outputs, factory-installed lock for security, and a pre-installed probe access port for external monitoring devices.
What construction and usability features support its healthcare application?
Includes a self‐closing glass door, hospital-grade “green-dot” power cord, adjustable wire shelving, LED interior lighting, front-breathing ventilation for built-in or freestanding use, sealed-back cabinet for ease of cleaning, and automatic defrost for maintenance convenience.
Shipping FAQ
What are the exterior dimensions and shipping weight?
Approximate external dimensions: 196 cm H × 59 cm W × 71 cm D (≈ 76.75″ × 23.38″ × 28.0″); shipping weight approx 80 kg (≈ 176 lb).
What electrical and installation requirements apply?
Operates on 115 V AC / 60 Hz, draws about 2.0 amps. Requires a dedicated grounded outlet, level installation surface, and sufficient clearance for ventilation (~51 mm / ~2″ behind unit) especially when built-in.
What steps should be taken during unpacking and placement?
On delivery inspect the unit for any visible shipping damage; if transported on its side, allow it to sit upright for several hours before powering; ensure it is level; maintain the required ventilation clearance and door-swing space; verify control and alarm functionality before loading contents.
What delivery or shipping method is typical for this model?
Given its specification and size this unit normally ships via palletized freight. Ensure receiving area access, clear path for installation, and stable off-loading surface prior to delivery.
Application FAQ
Is this unit suitable for vaccine or pharmaceutical cold-chain storage?
Yes — the ACR1602GNSF456LHD is explicitly designed for vaccine and clinical material storage, supporting the required +2 °C to +8 °C range and includes dedicated monitoring, alarms and security features aligned with healthcare standard practices.
What best practices should be followed for daily operation?
Use a calibrated data logger or min-max thermometer placed at mid-shelf height; label and rotate stored inventory regularly; avoid over-packing which may block airflow; minimize door-open time; and record internal temperature checks at start of each day as per facility SOP.
What routine maintenance is recommended?
Clean condenser/vent areas semi-annually; inspect and verify alarm, probe and thermostat functions monthly; check the door-gasket seal annually for integrity; keep the unit level and ensure ventilation paths are unobstructed.
What should be done in the event of a temperature excursion or alarm?
Document the time of the event, temperature deviation, and contents impacted; relocate items to approved backup storage if needed; identify and correct the root cause (door ajar, blocked airflow, power interruption); restore proper conditions and log corrective actions per facility protocol.