Description
ABS PH-ABT-HC-UCBI-0404G-ADA-LH Pharmacy Refrigerator Undercounter
The ABS PH-ABT-HC-UCBI-0404G-ADA-LH Pharmacy Refrigerator Undercounter feature a microprocessor temperature controller and display, which allows for precise temperature control, verification and recovery versus mechanical style thermostats.
The Built-In models are designed with a “front breathing” ventilation system, eliminating the need for clearance around the top and sides of the unit; enabling the user to incorporate these units into cabinets and other applications where free space is limited.
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.
Environmentally friendly, natural hydrocarbon (HC) refrigerants (not containing hydrofluorocarbons) vastly reduce global-warming potential (GWP) while meeting new EPA/SNAP mandates and UL, ASHRAE compliance guidelines.
Pharmacy Refrigerator Features:
- 4.6 cu. Ft. Capacity
- ADA compliant
- Left hinged door
- Three shelves (two adjustable/one fixed) with guard rail on back to prevent product from being pushed against the back wall
- Back wall evaporator cover
- Vaccine storage power cord warning label
- Forced draft circulation
- Keyed door lock
- Heavy duty door handle
- Magnetic door gasket for positive seal
- Leveling legs
- Led interior light
- Hfc-free refrigerant (r600a), foam insulation and packaging
- Pharmacy refrigerator/freezer toolkit and temperature logs
- Two year parts and labor warranty
Pharmacy Refrigerator Specifications:
- Exterior dimensions:
Width: 23 3/4″
Depth: 25 1/2”
Height: 31 15/16″ - Energy star certified
- Cycle defrost
- 115v, 60 hz, 1/8 hp, 1.74 amps
- Ul/c-ul listed
- Shipping weight: 150 lbs.
Pharmacy Refrigerator Temp Controller:
- Microprocessor temperature controller
- Adjustable operating temperature range: 36°f to 46°f (2°c to 8°c)
- Temperature monitor device (tmd) complies with the current CDC guidelines, with 3 years certification of calibration, “buffered” probe in the product simulated solution, min/max memory. F/c switchable, field installable, and visual & audible temp alarm
- Digital temperature display
- Audible and visual high and low temperature alarms
- Remote alarm contacts
- 2 probes; 1 in air and 1 in sample bottle
American Biotech Supply Refrigeration Solutions to Meet Tomorrow’s Standards. ABS Offers you creative product designs and an open mind for customer specific applications. View our wide selection of Premier Laboratory Refrigerators and Freezers.
We have a vast selection of laboratory refrigeration and freezer models to choose from, ranging down to Ultra-Low -86°C cascade freezers and -196°C cryogenic freezers.
Product FAQ
What does the ADA designation mean?
The ‘ADA’ designation indicates this unit is built to a lower height, meeting the Americans with Disabilities Act (ADA) requirements for accessibility in commercial and healthcare settings.
What is the significance of the LH designation?
The ‘LH’ indicates that the unit is configured with a Left-Hinged door, providing flexible placement options for specific pharmacy or lab layouts where door swing direction is critical.
What are the benefits of the glass door model?
The glass door allows for immediate visual identification of pharmaceuticals, minimizing the amount of time the door needs to remain open and thereby preserving internal temperature stability.
What makes this unit suitable for professional pharmacy use?
This unit is purpose-designed for pharmacy applications, featuring microprocessor-controlled temperature management and an integrated alarm system to ensure medication stability and regulatory compliance.
What is the temperature range?
The unit maintains a stable operating temperature range specifically suited for the storage of sensitive pharmaceuticals, typically between 2°C and 8°C.
What kind of warranty is included?
It comes with a standard two-year parts and labor warranty, plus an additional five-year compressor parts warranty to ensure long-term reliability.
Shipping FAQ
How is this undercounter unit shipped?
This medical-grade pharmacy refrigerator is shipped via dedicated LTL freight to ensure the refrigeration system arrives in optimal, stable condition.
Do I need liftgate service?
Yes, unless your facility is equipped with a loading dock, liftgate service is strongly recommended to safely lower the unit to ground level.
What should I inspect upon arrival?
Please carefully inspect the glass door, the unit, and all external packaging for any signs of transit impact or damage before signing the delivery receipt.
Any setup tips?
We recommend allowing the refrigerator to stand upright for at least 24 hours after arrival to allow compressor oils to settle fully before powering it on.