Description
Summit Vaccine Medical Freezer Description:
Accucold offers a quality selection of 20″ wide built-in undercounter all-freezers designed for installation under lower ADA compliant counters, with advanced features for the safe storage of select vaccines, pharmaceuticals, and other materials with strict storage conditions.
The Summit ADA305AFTBC Vaccine Medical Freezer is sized at 32″ high to fit under lower counters complying with ADA guidelines for accessibility. The fully finished textured white cabinet allows freestanding use, while the front-breathing system means this unit can be built-in with minimum clearance. The door is 3″ thick for efficient insulation, with a top-mounted lock for added security and convenient access. A digital thermostat, located in the kickplate, makes for easy temperature management with a readout in Fahrenheit or Celsius. This unit comes with a hospital grade cord with a green dot plug for added safety in high traffic areas. Powerful magnetic gaskets ensure a tight seal.
The Summit ADA305AFTBC Vaccine Medical Freezer includes a factory-installed towel bar style handle made from pure copper. Through its natural antimicrobial properties, copper will kill microbes, including viruses, in a shorter timeframe than most other metal surfaces allowing for a safer, cleaner user experience.
Inside, the Summit ADA305AFTBC Vaccine Medical Freezer can reach temperatures as low as -25°C. Static manual defrost operation provides efficient cooling inside the 2.47 cu.ft. interior. Three removable pull-out drawers on fixed wire shelving offer convenient storage, with plastic fronts that help to protect contents from warmer ambient temperatures when the door is opened. As recommended by the CDC, this unit has a flat door liner to ensure users don’t store sensitive contents on the door. The user-set high and low temperature alarms provide added security should the temperature go beyond the setpoint, while the open door alarm sounds if the door is ajar beyond one minute for added peace-of-mind. These alarms can be disabled depending on your needs. A preinstalled probe hole is included to allow users to connect their own monitoring equipment.
Sized at just under 20″ wide, the ADA305AFTBC makes an ideal choice for various facilities in need of a dependable freezer capable of -20ºC/-25ºC operation. For a complete set, pair this unit with the matching ADA404REFTBC all-refrigerator.
NOTE: Accucold freezers are purpose-built to the support meeting the current CDC (Centers for Disease Control) guidelines for the storage of vaccines. As referenced in the current Vaccine Storage and Handling Toolkit, these units incorporate several key features for stable operation. For additional temperature assurance, the CDC recommends a calibrated digital data logger (DDL), which can be added with the factory-installed option DL2B or addition of the DL2BKit accessory. Select Accucold products have been tested and certified to the new NSF 456 standard, and are now available for purchase. For additional information, contact our products specialists.
Summit Vaccine Medical Freezer Overview:
- ADA compliant design– 32″ height allows easy installation under lower ADA compliant counters
- -25°C capable– Low temperature operation ideal for vaccines, pharmaceuticals, and other sensitive materials
- Commercially approved– Meets ETL-S to NSF-7 standards for use in commercial settings
- Built-in capable– Front-breathing design lets you make the best use of space by installing your appliance under the counter
- Copper handle– Naturally antimicrobial, the towel bar handle made from pure copper helps to protect against microbes for added safety and cleanliness
- Factory-installed lock– Located towards the top of the door for easy access and to provide added security of unit (2 keys included)
- Access port– Unit comes with a factory installed probe hole for users to connect their own monitoring equipment
- Adjustable digital thermostat– Electronic controls located in the front kickplate for easy & institutive temperature management, with a readout in Fahrenheit or Celsius
- Basket shelves– Three slide-out baskets offer convenient storage, with plastic fronts to help protect contents from warmer air when the door is opened
- 3″ thick door– Improved insulation for an efficient performance; NOTE: due to the oversized design, the door does open in its own radius. If installing this freezer against a wall, please see our drawings to ensure adequate space
- Temperature alarms– User-set high/low temperature alarms provide added security if temperature goes beyond setpoint
- Open door alarm– Audible alarm will sound if the door is left ajar for over a minute
- Hospital grade cord with ‘green dot’ plug– Secure three-pronged cord for added unit safety
- Manual defrost– More temperature stability with an energy efficient manual defrost system
- Flat door liner– Interior design conforms to CDC guidelines to ensure stable conditions for sensitive materials
- No internal fans– Temperature remains more stable with a static cooling system
- Sealed back– Space-saving design that offers easier cleanability
- Magnetic door gasket– Quality door gaskets ensure a tight seal for energy savings and improved performance
- Matching all-refrigerator– This unit can be paired with the ADA404REFTBC all-refrigerator for a complete set
- Fully finished white cabinet– Allows the unit to be freestanding
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
- CARB Compliant– Environmentally friendly hydrocarbon-based refrigerant for a greener footprint that meets current CARB regulations
Summit Vaccine Medical Freezer Specification:
- Height of Cabinet– 31.75″
- Height to Hinge Cap– 32.0″
- Width– 19.75″
- Width with Door Open– 20.25″
- Depth– 22.5″
- Depth with Handle– 25.00″
- Depth with door at 90°– 40.25″
- Capacity– 2.47 cu.ft.
- Defrost Type– Manual
- Door– White
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Sanitation– ETL-S
- Amps– 1.1
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 65.0 lbs.
- Shipping Weight– 70.0 lbs.
- Parts & Labor Warranty– 1 Year
- Compressor Warranty– 5 Years
Summit Vaccine Medical Freezer Features:
- Door Swing– RHD
- Reversible– Yes
- Interior Height– 23.75″
- Interior Width– 14.5″
- Interior Depth– 17.0″
- Shelf Type– Drawers
- Shelf Qty– 3
- Thermostat Type– Digital
- Fan Type– Exterior
- Refrigerant Type– R600a
- Refrigerant Amount– 1.34 oz.
- High Side PSI– 180.0
- Low Side PSI– 80.0
- Level Legs– 2
- Interior Drawers– Yes
- Drawers Qty– 3
- Compressor Step Height– 5.75″
- Compressor Step Width– 14.5″
- Compressor Step Depth– 6.0″
- Temperature Range– -25° to -15°C
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the ADA305AFTBC and what is it used for?
The ADA305AFTBC is a 20-inch wide under-counter/ADA-compliant all-freezer (~2.47 cu ft / ~70 L) purpose-built for vaccine, pharmaceutical and medical storage. It reaches down to –25 °C, includes a pure copper antimicrobial handle, factory-installed lock, and is designed for built-in or freestanding use in clinical environments.
What are the key technical specifications?
• Capacity: 2.47 cu ft (~70 L)
• Dimensions: Height 81 cm (32″) × Width 50 cm (19.75″) × Depth 57 cm (22.5″) (door open ~102 cm)
• Temperature range: down to –25 °C (–25 °C to –15 °C for freezer zone)
• Defrost type: Manual
• Door & handle: White door/cabinet, antimicrobial copper towel-bar handle, factory-installed lock.
• Electrical: 115 V AC / 60 Hz, approx. 1.1 Amps
• Features: Front-breathing design (built-in capable), probe port for external monitoring, high/low temperature alarms, open-door alarm.
What safety and monitoring features are included?
– Visual & audible high/low temperature and open-door alarms.
– Factory-installed probe port for external monitoring equipment.
– Antimicrobial copper handle for added infection-control benefit.
– Self-locking door key mechanism, hospital-grade “green dot” plug cord.
Which environments is this unit best suited for?
Ideal for under-counter or built-in installations in hospitals, clinics, pharmacies, or immunization centers needing secure, compact, low-temperature storage for vaccines or other sensitive materials. Not intended for ultra-low (e.g., -80 °C) sample storage without further validation.
What warranty coverage is provided?
Parts & labor: 1 year; compressor parts: 5 years.
Shipping FAQ
How long does shipping take after placing an order?
Estimated shipping lead time: approximately 3-5 business days when in stock.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping is possible. For delivery to Melbourne, Australia consider:
• Import duties/customs clearance including Goods & Services Tax (GST).
• Site/doorway access: unit weight approx. 29 kg (~65 lbs) with palletized packaging; doorway width ≥50 cm, corridor & elevator clearance, floor loading.
• Electrical compatibility: This model is rated for 115 V AC / 60 Hz (US spec) — for Australia (230 V/50 Hz) ensure you order compatible variant or use a suitable transformer/adaptor.
How is the unit packaged for transit and what site requirements should be met?
The unit ships palletized/boxed; unpacked footprint ~81 cm H × ~50 cm W × ~57 cm D; allow ~102 cm depth clearance for door-open. Delivery site should have a firm, level surface and handling access (e.g., pallet jack or lift).
What should I do if my shipment arrives damaged or is delayed?
On delivery inspect the packaging and unit for visible damage or transit issues. If damage is found: photograph condition, keep packaging and delivery receipt, then contact your supplier with order/tracking details to initiate a freight claim or replacement.
Application FAQ
How do I install and set up the ADA305AFTBC?
1. Position the unit upright on a firm, level indoor surface (or under-counter location) with clearance behind and above for ventilation; ensure door-open clearance (~102 cm) is unobstructed.
2. Connect to a dedicated grounded outlet matching its specification (115 V AC / 60 Hz) or correct variant for your region (e.g., 230 V/50 Hz in Australia).
3. Plug-in and power the unit. Set the digital thermostat to your target temperature (recommended -25 °C for vaccine freeze storage) and allow several hours for internal temperature to stabilize before loading contents.
4. Verify the door lock, copper handle operation, alarms (door-open, temp high/low), probe port functionality and shelving draw-outs are functioning correctly.
What best practices should I follow when using this vaccine freezer?
• Load items evenly in the three pull-out drawers; avoid storing on the door or blocking drawers to maintain uniform cooling.
• Minimise door-open time; use the glass-free solid door insulation and rely on external monitoring when possible.
• Label stored materials (date/time, batch) and rotate stock (first-in/first-out) especially if used for vaccine programs.
• Keep the interior and door gasket clean; ensure unit remains level to maintain performance and seal integrity.
What should I do if a temperature excursion or power interruption occurs?
If internal temperature rises above or falls below your target, or a power interruption occurs:
• Record the event: date/time, observed temperature, contents impacted.
• Relocate critical contents to validated backup storage if required.
• Investigate the root cause (door left open, ambient heat load, power or sensor fault).
• After normal operating temperature is restored and stable, verify stored contents before resuming regular use.
What maintenance and verification routines are recommended?
• Monthly: Inspect door seal/gasket, handle/lock mechanics and ensure the unit remains level.
• Quarterly: Clean the exterior cabinet and interior drawers, verify thermostat operation, alarms and probe port, remove any dust/debris from ventilation pathways.
• Periodically: Maintain internal logs of minimum/maximum temperatures and service/maintenance records for audit compliance in medical/vaccine storage settings.