Description
Summit Vaccine Refrigerator Freezer Description:
Accucold offers innovative solutions to vaccine storage with stackable purpose-built refrigerator-freezer combinations design to support meeting CDC/VFC guidelines.
Sized just under 24″ wide, the ARS6PV-VT65MLSTACKMED2 combines the 6 cu.ft. ARS6PV auto defrost all-refrigerator with VT65MLBIMED2 all-freezer into one slim footprint using Accucold’s StackRack. This ships in three boxes and requires some minor assembly upon delivery.
Refrigerator Summit ARS6PV-VT65MLSTACKMED2 Vaccine Refrigerator Freezer is part of Accucold’s Pharma-Vac Performance Series, designed to meet the demanding needs of vaccine storage. It utilizes a microprocessor temperature controller with an adjustable control range from +2 to +8ºC. The buffered temperature probe is encased in a glycol-filled bottle to better simulate temperature of the stored product, with an additional sensor that reads air temperature to ensure superior temperature control. The current and min/max temperature display is viewable in Celsius or Fahrenheit to the nearest tenth of a degree. The audible and visual temperature alarm sounds if the unit’s interior goes out of range. Additional alarms include a power failure alert, door open alarm, and sensor failure alarm. This unit is equipped with remote alarm contacts in the rear and a 1/2″ probe port on the right side to accept additional monitoring devices. A factory-installed antimicrobial handle is powder-coated with naturally occurring silver ions, which help to reduce the spread of germs, allowing for a safer, cleaner user experience. This unit comes with a hospital grade cord with a green dot plug for added safety in high traffic areas.
Inside, the Summit ARS6PV-VT65MLSTACKMED2 Vaccine Refrigerator Freezer utilizes adjustable cycle defrost operation to avoid temperature spikes. Optimized forced air cooling with an internal fan ensures excellent temperature stability and uniformity with rapid recovery after door openings. The adjustable plastic-coated wire shelves can be positioned at 1/2″ increments to accommodate virtually any sized item. This unit includes LED internal lighting with an on/off rocker switch. The interior is constructed from white powder coated metal to ensure added durability and better temperature retention.
The refrigerator is designed to be stacked on top of the VT65MLBIMED2, a -25ºC all-freezer that utilizes manual defrost operation. The freezer includes three slide-out plastic storage drawers to better protect the contents from warmer ambient temperature when the door is opened. The digital control panel includes a NIST calibrated thermometer that provides a continuous display of the current and high/low temperature. This unit is also equipped with a temperature alarm and access port to accept additional monitoring equipment.
Both units include front-mounted keyed locks and powerful magnetic gaskets to ensure a positive seal. The refrigerator has a self-closing door for added protection in high traffic settings. Additional stacked combinations are available. Browse Accucold’s website for more information, or contact our sales team at 718-893-3900 to discuss your equipment needs.
NOTE: Accucold refrigerators are purpose-built to the support meeting the current CDC (Centers for Disease Control) guidelines for the storage of vaccines. At this time, Accucold refrigerators have not been evaluated to the new NSF 456 standard. As referenced in the current Vaccine Storage and Handling Toolkit, these units incorporate microprocessor controls and forced air circulation for stable operation between 2 and 8ºC. For additional temperature assurance, the CDC recommends a calibrated digital data logger (DDL), which can be added with the factory-installed option DL2B or addition of the DL2BKit accessory.
Summit Vaccine Refrigerator Freezer Features:
- Space-saving design– Allows you to create a full refrigerator-freezer with independent controls in a slim-fitting footprint
- Easy assembly– Unit ships in three boxes for customer to install the StackRack on the bottom refrigerator/freezer and place and secure the other unit on top
- Purpose-built design– Designed and purpose-built for pharmacy, medication, and vaccine applications to support meeting CDC/VFC guidelines
- +/-1ºC Variation (refrigerator)– The published temperature variation is derived from the maximum deviation of an NTC sensor in a 1 oz. vial located nearest the chamber geometric center during a 24 hour test period
- Adjustable temperature– Intuitive controls let you adjust the interior temperature between +2 to +8ºC (refrigerator) and -20º to -15ºC (freezer)
- Temperature alarms– Visual and audible alarms alert if the interior temperature rises or falls out of set range
- Remote alarm contacts (refrigerator)– Electronic controller is equipped with a volt-free contact for alarm forwarding to an external remote system (contacts located in the external rear of the unit)
- Automatic defrost with adjustable cycle defrost (refrigerator)– Adjustable cycle defrost operation ensures the unit avoid temperature spikes and ice buildup
- Manual defrost (freezer)– Static cooling system for reduced temperature fluctuations
- Access ports– Both units include a pre-installed probe hole to allow easier connection to user-provided monitoring equipment
- Open door alarm (refrigerator)– Alarm will sound if the door is left ajar for over 4 minutes (adjustable from 1-99 minutes)
- Audible power failure alarm (refrigerator)– Battery-powered alarm sounds if the power connection is interrupted
- Sensor failure alarm (refrigerator)– Audible and visual alarm warns if the sensor is disrupted
- Password protection (refrigerator)– Control panel requires password entry to adjust the parameters beyond the set point
- Microprocessor control panels– Digital controls with a large LED readout in Celsius or Fahrenheit for intuitive temperature management
- Resettable min/max display (refrigerator)– Control panel provides a readout of the low and high temperatures in 15 minute increments
- Factory-installed locks– Keyed lock conveniently located towards the top of the door for added security (2 keys included for each unit)
- Adjustable plastic-coated wire shelves for improved circulation– Designed to accommodate a variety of sizes for flexible storage
- Interior light (refrigerator)– Energy efficient LED lighting with internal on/off switch
- Self-closing door (refrigerator)– Added protection in busy settings; NOTE: door is not self-closing once reversed to an LHD swing
- Magnetic door gaskets– Quality door gaskets ensure a tight seal for energy savings and improved performance
- Extended warranty– 2 years parts and labor/5 years compressor
- Eco-Friendly (refrigerator)– Using environmentally friendly hydrocarbons refrigerants for a greener footprint
- Usage stickers– Ships with removable warning stickers for “DO NOT UNPLUG,” “DO NOT ADJUST TEMPERATURE,” and “DO NOT STORE PERSONAL ITEMS”
- Versatile design (refrigerator)– Compatible with Pyxis®, Omnicell®, and AcuDose RX®
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
Summit Vaccine Refrigerator Freezer Overview:
- Height of Cabinet– 68.0″
- Width– 23.63″
- Depth– 24.38″
- Depth with Handle– 26.13″
- Capacity– 9.2 cu.ft.
- Door– White
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Voltage/Frequency– 115 V AC/60 Hz
- Estimated Time to Ship– 5 to 7 days
- Parts & Labor Warranty– 2 Years
- Compressor Warranty– 5 Years
Summit Vaccine Refrigerator Freezer Features:
- Door Swing– RHD
- Reversible– Factory Reversible
- Adjustable Shelves– Yes
- Interior Light– Yes
- Refrigerator – Interior Height– 24.5″
- Refrigerator – Interior Width– 20.0″
- Refrigerator – Interior Depth– 17.0″
- Refrigerator – Capacity– 6.0 cu.ft
- Refrigerator – Shelf Type– Wire
- Refrigerator – Shelf Qty– 4
- Refrigerator – Temperature Range– 2 to 8ºC
- Refrigerator – Refrigerant Type– R600a
- Refrigerator – Refrigerant Amount– 1.66 oz.
- Freezer – Interior Height– 24.25″
- Freezer – Interior Width– 17.63″
- Freezer – Interior Depth– 17.0″
- Freezer – Capacity– 3.2 cu.ft
- Thermostat Type– Digital
- Fan Type– Interior
- Freezer – Temperature Range– -12º to -25ºC
- Freezer – Refrigerant Type– R134a
- Freezer – Refrigerant Amount– 2.76 oz.
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is this unit and what is it used for?
The ARS6PV-VT65MLSTACKMED2 is a stacked refrigerator/freezer combination purpose-built for pharmacy, medication and vaccine storage applications. It pairs an ~6 cu ft (~170 L) +2 °C-to-+8 °C refrigerator section with an ~3.2 cu ft (~90 L) freezer section (-12 °C to -25 °C) in one slim footprint (~60 cm wide).
What are the key technical specifications?
• Capacity: 9.2 cu ft (~261 L) combined.
• Dimensions: Height ~173 cm, Width ~60 cm, Depth ~62 cm (depth with handle ~66 cm).
• Refrigerator section: ~6.0 cu ft; temperature range ~+2 °C to +8 °C.
• Freezer section: ~3.2 cu ft; temperature range approx. -12 °C to -25 °C.
• Ships as a stack of three boxes (bottom refrigerator/freezer plus StackRack plus top unit).
What safety and monitoring features are included?
Features include: digital thermostat/microprocessor control, factory-installed probe ports for external monitoring, visual & audible high/low temperature alarms, open and power-failure alarms (on the refrigerator section), factory lock on door, antimicrobial silver-ion coated handle on refrigerator section.
Which environments is this unit best suited for?
Ideal for clinics, vaccination centers, pharmacies or any medical facility requiring both standard refrigeration (+2 °C to +8 °C) and freezer storage (< -12 °C) in a compact footprint. Not designed for ultra-low temperature (< -80 °C) applications unless otherwise specified.
What warranty coverage is provided?
Manufacturer specification: 2 years parts & labor; 5 years compressor parts.
Shipping FAQ
How long does shipping take after placing an order?
Estimated shipping time when in stock: 5-7 business days.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping can be arranged. When shipping to Melbourne, Australia you should consider:
• Import duties, customs and GST.
• Site access: width ~60 cm, depth ~62 cm (plus ~66 cm with handle) clearance; allow additional depth for door-open swing.
• Electrical compatibility: this unit is rated at 115 V AC/60 Hz (US spec). For Australia (230 V/50 Hz) ensure correct variant or arrange suitable reform.
How is the unit packaged for transit and what site requirements should be met?
The unit ships in three parts (stacked configuration). Unpacked footprint about 173 cm H × 60 cm W × 62 cm D; door swing clearance should be confirmed. Delivery site should have a level, firm surface and clear access from delivery to installation point.
What should I do if my shipment arrives damaged or is delayed?
Upon delivery inspect packaging and unit for signs of transit damage or delay. If damage is present: photograph condition, retain all packaging and delivery documentation, then contact your supplier with order/tracking details to initiate a freight claim or arrange replacement.
Application FAQ
How do I install and set up the ARS6PV-VT65MLSTACKMED2?
1. Move all three boxes into the installation area, assemble the StackRack per manufacturer instructions so the refrigerator is at the bottom, then secure the freezer section on top.
2. Position the unit upright on a firm, level indoor surface; ensure clearance behind and above for ventilation and allow full door opening (~66 cm depth).
3. Connect the unit to a dedicated grounded outlet matching its specification (115 V AC/60 Hz for the US version) or the correct local variant for your region (Australian 230 V/50 Hz) and power on.
4. Set the digital controller for the refrigerator section to target range (+2 °C to +8 °C) and for the freezer section to target range (-12 °C to -25 °C); allow several hours (typically 4-6 hrs) for temperature stabilization before loading sensitive items.
What best practices should I follow when using this refrigerator/freezer combination?
• Use the adjustable wire shelves and keep space around items for airflow; avoid over-packing either the fridge or freezer section.
• Label stored items (date/time, batch) and use First-In/First-Out (FIFO) stock rotation when applicable.
• Minimize door-open time: rely on external display/monitoring instead of frequent door checks.
• Regularly inspect the door gasket/seal, door self-close mechanism (if included), the lock, and cleaning of interior surfaces.
What should I do if a temperature excursion or power interruption occurs?
If interior temperature moves outside the target range or power interruption occurs:
• Record the event (date/time, observed temperature, contents impacted).
• Move critical contents to validated backup storage until the system returns to stable operation.
• Investigate the root cause (door left open too long, ambient heat load, power fault or sensor failure).
• Once conditions stabilize and temperature returns within range, verify contents before returning to routine storage.
What maintenance and verification routines are recommended?
• Monthly: Check door gasket/seal condition, hinges/handle/lock mechanism, and ensure the unit remains level on its surface.
• Quarterly: Clean external surfaces, inspect ventilation inlet/outlet (if applicable), verify alarm and probe-port functionality.
• Periodically: Inspect and record minimum/maximum internal temperature logs, maintain calibration certificates for any data logger equipment, and archive service records — particularly important for vaccine/medical-grade compliance.