Description
Summit Undercounter Medical Vaccine Refrigerator Description:
The Summit FF7LWBIVACADA Undercounter Medical Vaccine Refrigerator are specially designed to store sensitive medical and pharmaceutical materials under the proper conditions, with innovative construction for temperature stability and user-friendly features for added convenience.
Sized with a 24″ footprint and 32″ height, the Summit FF7LWBIVACADA Undercounter Medical Vaccine Refrigerator is an undercounter all-refrigerator that can be used built-in or freestanding in ADA compliant settings. The automatic defrost interior is equipped with fan-forced cooling to ensure even temperature distribution from top to bottom. Three slide-out wire basket drawers make it easy to organize the interior while the plastic drawer fronts help to shield stored contents from being exposed to room temperature when the door is open. As recommended by the CDC for vaccine storage, this unit has a flat interior door liner.
The Summit FF7LWBIVACADA Undercounter Medical Vaccine Refrigerator includes a factory installed front lock for added security. The self-closing door is an ideal feature for lab or medical settings, helping to protect valuable samples and vaccines when a busy staff may not always have time to ensure a tight seal. The hospital grade ‘green dot’ cord offers added safety in high-traffic locations.
For temperature precision, the Summit FF7LWBIVACADA Undercounter Medical Vaccine Refrigerator includes a digital thermostat. A high/low temperature alarm sends an audible alert if the unit is out of set range, while also displaying a NIST calibrated temperature readout to the nearest tenth of a degree. An access port with a contact is pre-installed, allowing users to install additional monitoring equipment. Customers can specify the access port location upon order.
In addition to the standard features of the VAC line, the Summit FF7LWBIVACADA Undercounter Medical Vaccine Refrigerator can be further customized to your exact specifications. Additional options include temperature loggers, keypad locks, open door alarms, and more. For more information, call our office to discuss your complete medical refrigeration needs with our sales team.
Summit Undercounter Medical Vaccine Refrigerator Features:
- Improved temperature stability– Interior is designed to operate within 2ºC of set temperature performance
- ADA Compliant– 32″ height is designed to fit under lower ADA compliant counters
- Interior basket drawers– Three slide-out wire baskets with plastic fronts offer convenient storage solutions while protecting stored items when the door is open
- Digital thermostat– Electronic controls ensure precise and intuitive temperature management
- High/low temperature alarm– Audible alarm sounds if the interior temperature rises or falls out of set range
- NIST calibrated temperature display– Alarm includes NIST calibrated display of the current interior and high/low temperature
- Self-closing door– Door closes automatically when opened at angles less than 90º
- Access port with dry contact– Pre-installed probe with dry contact allows users to install their own monitoring equipment; customers can specify placement of the access port upon order
- Hospital grade cord and plug– Secure three-pronged ‘green dot’ cord for added unit safety in high traffic settings
- Automatic defrost– Reduced user maintenance with automatic defrost
- Internal fan– Fan-cooled interior maintains a consistent temperature from top to bottom
- Factory installed lock– Keyed lock for a secure interior
- Flat door liner– Interior design conforms to CDC guidelines to ensure stable conditions for sensitive materials
- Built-in capable– Front-breathing design allows built-in installation under counters
- Fully finished white cabinet– Allows the unit to be used freestanding
- 100% CFC free– Environmentally friendly design does not contain ozone-damaging chemicals
Summit Undercounter Medical Vaccine Refrigerator Specifications:
- Height of Cabinet– 32.0″
- Height to Hinge Cap– 32.25″
- Width– 23.63″
- Depth– 23.25″
- Depth with Handle– 25.13″
- Depth with door at 90°– 44.25″
- Capacity– 2.6 cu.ft.
- Defrost Type– Automatic
- Door– White
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Amps– 1.4
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 100.0 lbs.
- Shipping Weight– 105.0 lbs.
- Parts & Labor Warranty– 2 Year
- Compressor Warranty– 5 Years
Summit Undercounter Medical Vaccine Refrigerator Features:
- Door Swing– RHD
- Reversible– Factory Reversible
- Thermostat Type- Digital
- Fan Type– Interior and Exterior
- Refrigerant Type– R600a
- Refrigerant Amount– 0.64oz.
- High Side PSI– 180.0
- Low Side PSI– 80.0
- Level Legs Qty– 4
- Interior Drawers– Yes
- Interior Drawers Qty– 3
- Temperature Range– 2º to 8ºC
Summit Undercounter Medical Vaccine Refrigerator Dimensions:
- Interior Height– 22.5″
- Interior Width– 17.25″
- Interior Depth– 15.0″
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the FF7LWBIVACADA and what is it used for?
The FF7LWBIVACADA is a 2.6 cu ft (~74 L) under-counter built-in all-refrigerator (ADA-height ~81 cm) purpose-designed for vaccine, medication and pharmacy-grade storage environments. It maintains a controlled interior temperature of +2 °C to +8 °C and includes features tailored for sensitive medical storage.
What are the key technical specifications?
• Capacity: 2.6 cu ft (~74 L).
• External dimensions: Height ~81 cm (32.0″) H × Width ~60 cm (23.63″) W × Depth ~59 cm (23.25″) D; door-open depth approx. 112 cm (44.25″)
• Temperature range: +2 °C to +8 °C.
• Defrost type: Automatic, fan-assisted interior for uniform cooling.
• Shelving/Interior: 3 slide-out wire basket drawers with plastic fronts (to shield contents when door opens).
• Construction/Features: Factory-installed keyed lock; self-closing door; flat interior door liner as recommended for vaccine storage.
• Power (U.S. version): 115 V AC / 60 Hz; approx. 1.4 amps.
What safety and security features are included?
Features include:
– Keyed door lock for secure access.
– High/low temperature audible alarm plus NIST-calibrated display of current and high/low temps.
– Access probe port (3/8″) pre-installed for external monitoring equipment.
– Magnetic door gasket, foamed insulation, and hospital-grade “green dot” power cord for reliability.
Which environments is this unit best suited for?
Ideal for pharmacies, immunization clinics, medical-storage rooms or vaccine-storage environments where compact, under-counter +2 °C to +8 °C refrigeration is required in a footprint that is ADA-compliant under counters. Not intended for ultra-low temperature (-20 °C or colder) or freezer applications unless separately validated.
Shipping FAQ
How long does shipping take after placing an order?
Estimated lead time: 5-7 business days when the unit is in stock.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping is available. For delivery to Melbourne, Australia you should consider:
• Import duties, customs clearance and GST.
• Site access: ensure doorways/elevators/corridor width can accommodate ~60 cm width × ~59 cm depth and allow approx. 112 cm front clearance for full door swing.
• Electrical compatibility: This model is rated for 115 V AC / 60 Hz (U.S. specification) — for Australia (230 V / 50 Hz), order the correct variant or use a certified step-down transformer by a licensed electrician.
How is the unit packaged for transit and what site requirements should be met?
The unit ships boxed and palletized; unpacked footprint approx. Height ~81 cm × Width ~60 cm × Depth ~59 cm; door-open depth approx. 112 cm. Delivery path must be level, firm and free of obstructions from vehicle to installation point.
What should I do if my shipment arrives damaged or delayed?
On delivery inspect the outer packaging and unit for visible damage. Photograph any damage, retain packing materials and delivery receipt, then contact your supplier with order/tracking details to initiate a freight claim or arrange replacement.
Application FAQ
How do I install and set up the FF7LWBIVACADA?
1. Position the unit under a counter or on a firm, level indoor surface in a climate-controlled room; allow sufficient ventilation at the rear and top.
2. Ensure front clearance of approx. 112 cm (door open) is available and width ~60 cm to fit the unit.
3. Connect to a dedicated grounded outlet matching the unit’s voltage variant.
4. Power on the unit, set the digital thermostat to between +2 °C and +8 °C and allow approximately 4–6 hours for internal temperature stabilization before loading sensitive vaccines or medication.
What best practices should I follow when using this refrigerator?
• Store items in the three slide-out basket drawers with adequate spacing for airflow; avoid over-packing to maintain uniform cooling.
• Label stored items (batch number, date/time) and apply first-in/first-out (FIFO) rotation where applicable.
• Minimize door-open duration; use the digital display and external monitoring port rather than frequently opening the unit.
• Regularly inspect door gasket/seal, hinges, lock mechanism and ensure the unit remains level. Clean interior surfaces when necessary.
What should I do if a temperature excursion or power interruption occurs?
If internal temperature rises above or falls below the +2 °C to +8 °C range or a power interruption occurs:
• Record the event: date/time, observed temperature deviation and impacted contents.
• Transfer critical contents to validated backup storage until conditions stabilize.
• Investigate root cause (door left open, high ambient load, power/sensor fault).
• Once the unit returns to target temperature and stabilizes, inspect contents for integrity before resuming normal operation.
What maintenance and verification routines are recommended?
• Monthly: Inspect door gasket/seal, hinges/lock mechanism and check level of unit.
• Quarterly: Clean interior and exterior surfaces; verify digital thermostat/alarm functions, probe port accessibility and fan operation.
• Periodically: Review any logged temperature data if monitoring enabled, verify calibration of monitoring equipment and keep documented maintenance/service records for compliance.