Description
Summit Compact Vaccine Medical Freezer Description:
The Summit FS30LVAC Compact Vaccine Medical Freezer Accucold® VAC Series are specially designed to store sensitive medical and pharmaceutical materials under the proper conditions, with innovative construction for temperature stability and user-friendly features for added convenience.
Summit FS30LVAC Compact Vaccine Medical Freezer is a compact all-freezer designed to operate at -20º C, the ideal temperature for long-term storage of select sensitive materials. The 1.8 cu.ft. manual defrost interior offers static cooling for stable results. Removable wire shelving allows easy storage options, with a seamless interior liner to simplify cleanup. This unit has a flat door liner as recommended by the CDC.
For optimum temperature assurance, the Summit FS30LVAC Compact Vaccine Medical Freezer includes a digital thermostat and a high/low temperature alarm with a NIST calibrated temperature display to the nearest tenth of a degree. A hospital grade cord includes a ‘green dot’ plug for added safety in high traffic settings, while the self-closing door ensures stored items aren’t accidentally exposed to room temperature. This model comes in a white finish and includes a factory installed lock. An access port with a contact is pre-installed, allowing users to install additional monitoring equipment. Customers can specify the access port location upon order.
The Summit FS30LVAC Compact Vaccine Medical Freezer can be paired with the FF28LWHVAC all-refrigerator for a matching set. These units can also be ordered with more advanced medical features, including a power-out alarm, temperature logger, and more.
Summit Compact Vaccine Medical Freezer Features:
- Compact dimensions– Sized for easy use on top of counters
- Digital thermostat– Electronic controls ensure easy and intuitive temperature management, with a readout in Celsius or Fahrenheit
- High/low temperature alarm– Audible alarm sounds if the interior temperature rises or falls out of set range
- NIST calibrated temperature display– Alarm includes NIST calibrated display of the current interior and high/low temperature to the nearest tenth of a degree
- Self-closing door– Door closes automatically when opened at angles less than 90º
- Access port with dry contact– Pre-installed probe with dry contact allows users to install their own monitoring equipment; customers can specify placement of the access port upon order
- Hospital grade cord with ‘green dot’ plug– Secure three-pronged cord for added unit safety
- Manual defrost– Static cooling system for improved temperature stability
- Factory installed lock– Keyed lock for a secure interior
- Removable shelves– Wire shelves can be removed to accommodate different sizes or for easy cleanup
- Flat door liner– Interior design conforms to CDC guidelines to ensure stable conditions for sensitive materials
- One piece interior liner– Easy clean-up with a seamless liner that won’t hold a mess in hidden crevices
- Factory reversible door– Standard unit ships RHD but can be ordered LHD
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
Summit Compact Vaccine Medical Freezer Specifications:
- Height of Cabinet– 26.25″
- Height to Hinge Cap– 24.63″
- Width– 18.5″
- Depth– 17.75″
- Depth with door at 90°– 34.5″
- Capacity– 1.8 cu.ft.
- Defrost Type– Manual
- Door– White
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Amps– 0.8
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 50.0 lbs.
- Shipping Weight– 60.0 lbs.
- Parts & Labor Warranty– 1 Year
- Compressor Warranty– 5 Years
Summit Compact Vaccine Medical Freezer Features:
- Door Swing– RHD
- Reversible– Factory Reversible
- Interior Height– 18.63″
- Interior Width– 13.38″
- Interior Depth– 13.0″
- Shelf Type– Wire
- Shelf Qty– 2
- Thermostat Type– Digital
- Refrigerant Type– R600a
- Refrigerant Amount– 0.8 oz.
- Level Legs– 2
- Compressor Step Height– 8.0″
- Compressor Step Width– 13.38″
- Compressor Step Depth– 5.75″
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the FS30LVAC and what is it used for?
The FS30LVAC is a 1.8 cu ft (~51 L) countertop all-freezer designed for vaccine, pharmacy, or medical storage environments. It maintains a stable low-temperature environment (approx. –20 °C to –15 °C) and includes features such as a keyed lock, self-closing door, digital thermostat, high/low alarm and probe access port for monitoring.
What are the key technical specifications?
• Capacity: 1.8 cu ft (~51 L)
• External dimensions: Height ~63 cm, Width ~47 cm, Depth ~45 cm; door-open depth approx. 88 cm.
• Temperature range: –20 °C to –15 °C.
• Defrost type: Manual.
• Shelves: 2 adjustable wire shelves.
• Door & security: Solid white door, factory-installed keyed lock, hospital-grade cord with ‘green dot’.
• Monitoring: Digital thermostat with high/low temperature alarm, NIST-calibrated display and 3/8″ access port for external probe.
• Electrical: 115 V AC / 60 Hz (U.S. version).
What safety and security features are included?
Features include:
– Keyed lock for secure access.
– Audible/visual high/low temperature alarm with NIST-calibrated digital read-out for accurate tracking.
– Self-closing door ensures stored contents are not left exposed to ambient conditions.
– Magnetic door gasket and flat door liner compliant with CDC storage guidelines.
– Access port enabling external monitoring/data-logging for regulated storage.
Which environments is this unit best suited for?
Ideal for vaccine clinics, immunization rooms, pharmacy cold-rooms, or medical-storage zones where compact freezer footprint and stable sub-zero temperature control (~–20 °C) are required. Not suited for ultra-low (≤ –80 °C) or large-volume freezer applications.
Shipping FAQ
How long does shipping take after placing an order?
Typical lead time: 5-7 business days once the unit is in stock and ready for dispatch.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping is available. For delivery to Melbourne, Australia you should consider:
• Import duties, customs clearance, and GST.
• Site access: ensure pathways/elevators/doorways accommodate approx. 47 cm width × 45 cm depth and allow ~88 cm front clearance for door-open.
• Electrical compatibility: The standard U.S. version runs at 115 V AC / 60 Hz — for Australia (230 V / 50 Hz), you must ensure you order the export variant or install a certified step-down transformer by a licensed electrician.
How is the unit packaged for transit and what site requirements should be met?
The unit ships boxed and palletized; unpacked footprint approx. 63 cm H × 47 cm W × 45 cm D; door-open depth approx. 88 cm. Ensure the delivery pathway is level, firm, and clear from the vehicle to the installation site.
What should I do if my shipment arrives damaged or delayed?
On delivery inspect the packaging and unit for visible transport damage. Photograph any damage, retain packaging and delivery receipt, and contact your supplier promptly with your order/tracking details to initiate a freight claim or arrange replacement.
Application FAQ
How do I install and set up the FS30LVAC?
1. Place the unit upright on a firm, level indoor surface in a climate-controlled environment; allow clearance behind/above and approx. 88 cm front clearance for full door opening.
2. Connect to a dedicated grounded outlet matching your variant’s voltage.
3. Power the unit, set the controller to the target –20 °C to –15 °C range, and allow 2-4 hours for internal temperature stabilization before loading sensitive contents.
What best practices should I follow when using this freezer?
• Place items on the adjustable wire shelves with spacing to allow airflow; avoid over-packing which can impede temperature uniformity.
• Label stored items with batch number/date/time and apply first-in/first-out (FIFO) rotation where applicable.
• Minimize door-open duration; rely on the digital display, alarm and monitoring port rather than frequent door access.
• Inspect door gasket/seal, hinges and lock mechanism regularly; ensure unit remains level and interior surfaces are clean and frost-free.
What should I do if a temperature excursion or power interruption occurs?
If internal temperature rises above or falls below the target –20 °C to –15 °C range, or a power interruption occurs:
• Record the event: date/time, observed temperature and impacted contents.
• Transfer critical items to validated backup freezer storage until proper conditions are restored.
• Investigate root cause (e.g., door left open, ambient heat load spike, power/sensor fault).
• Once internal temperature returns within the target range and stabilizes, inspect contents for integrity prior to resuming normal use.
What maintenance and verification routines are recommended?
• Monthly: Check door gasket/seal, hinges/lock; confirm the unit remains level.
• Quarterly: Clean interior/exterior surfaces; verify alarm and probe port functionality, and confirm insulation performance.