Summit Undercounter Medical Freezer Description:
Accucold by Summit Appliance’s MED series all-freezers come fully featured for stabilized temperature performance for use in medical, scientific, pharmaceutical, and laboratory institutions.
The Summit VT65MLMEDADA Undercounter Medical Freezer has a 24″ footprint and 32″ height, allowing it to be used in settings following ADA guidelines for lower counter height. It comes in a white finish and includes a front-mounted lock for added security. This unit is designed for freestanding use.
The manual defrost Summit VT65MLMEDADA Undercounter Medical Freezer operates several degrees colder than most freezers for safe long-term storage of food, vaccines, or pharmaceuticals. Temperatures can reach and maintain -25º C. Three removable pull-out basket drawers provide convenient storage solutions, while extra thick insulation limits energy waste and keeps the cold inside.
We include an audible NIST calibrated temperature alarm complete with an external display of the current and high/low temperature mounted on the door. A green dot hospital grade plug is installed for added safety in high traffic areas. An access port allows users to connect their own monitoring equipment.
With its slim size and reliable construction, the VT65MLMEDADA works well in any setting in need trusted storage. For a built-in version of this model, see the VT65MLBIMEDADA. Additionally, we offer a variety of options to customize this unit, including USB data loggers, digital thermostats, self-closing doors, and more.
NOTE: Accucold freezers are purpose-built to the support meeting the current CDC (Centers for Disease Control) guidelines for the storage of vaccines. As referenced in the current Vaccine Storage and Handling Toolkit, these units incorporate several key features for stable operation. For additional temperature assurance, the CDC recommends a calibrated digital data logger (DDL), which can be added with the factory-installed option DL2B or addition of the DL2BKit accessory. Select Accucold products have been tested and certified to the new NSF 456 standard, and are now available for purchase. For additional information, contact our products specialists.
Summit Undercounter Medical Freezer Features:
- ADA compliant–32 high meets ADA guidelines
- Factory installed lock–Keyed lock for secure interior
- -25ºC capable–Low temperature capability ideal for certain sensitive items
- High temperature alarm–Warns of temperature fluctuations and includes an external temperature readout on the door
- Manual defrost–Efficient performance with reliable temperature stability
- Hospital grade cord with ‘green dot’ plug–Secure three-pronged cord for added unit safety
- Pull-out drawers–Clear baskets for convenient and separated storage
- Access port– Unit comes with a factory installed probe hole for users to connect their own monitoring equipment
- Digital thermostat– Electronic controls located in the front kickplate for easy and intuitive temperature management
- Fully finished white cabinet–Allows the unit to be used freestanding
- 100% CFC Free–Environmentally friendly design with no ozone-damaging chemicals
Summit Undercounter Medical Freezer Specifications:
- Height of Cabinet– 32.0″
- Height to Hinge Cap– 32.25″
- Width– 23.63″
- Depth– 23.5″
- Depth with Handle–24.25″
- Depth with door at 90°– 45.0″
- Capacity– 3.2 cu.ft.
- Defrost Type– Manual
- Door– White
- Cabinet– White
- US Electrical Safety– UL
- Canadian Electrical Safety– ULC
- Amps– 1.3
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 105.0 lbs.
- Shipping Weight– 120.0 lbs.
- Parts & Labor Warranty– 1 Year
- Compressor Warranty– 5 Years
Summit Undercounter Medical Freezer Features:
- Door Swing– RHD
- Reversible– Factory Reversible
- Interior Height– 24.25″
- Interior Width– 17.75″
- Interior Depth– 16.5″
- Thermostat Type– Digital
- Refrigerant Type– R134a
- Refrigerant Amount– 2.7 oz.
- High Side PSI– 235.0
- Low Side PSI– 70.0
- Interior Drawers– Yes
- Drawers Qty– 3
- Compressor Step Height– 6.5″
- Compressor Step Width– 17.5″
- Compressor Step Depth– 5.5″
- Temperature Range– -12ºC to -25ºC
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the VT65MLMEDADA and what is it used for?
The VT65MLMEDADA is a 3.2 cu ft (~91 L) under-counter, ADA-compliant medical freezer designed for secure storage of temperature-sensitive materials such as pharmaceuticals or vaccines. It features manual defrost and can reach approximately –25 °C for reliable low-temperature storage.
What are the key technical specifications?
• Capacity: 3.2 cu ft (~91 L)
• External dimensions: Height ~81 cm, Width ~60 cm, Depth ~60 cm; door-open depth ~114 cm
• Temperature range: Approx. –12 °C to –25 °C
• Defrost type: Manual
• Interior: Three clear pull-out drawers for organized storage
• Security/monitoring: Factory-installed lock; digital thermostat; high/low alarm; access port for monitoring probe
• Electrical (US version): 115 V AC / 60 Hz; draws ~1.3 amps
What safety and security features are included?
Features include:
– Keyed door lock for controlled access
– Audible high/low temperature alarm
– Access port for external monitoring equipment
– Magnetic door gasket, extra thick insulation for stable low-temperature performance
Which environments is this unit best suited for?
Ideal for clinics, laboratories, and medical storage rooms where compact under-counter low-temperature freezer capacity (~ –25 °C) is needed in a restricted footprint. Not intended for ultra-low (≤ –80 °C) applications unless specifically validated.
Shipping FAQ
How long does shipping take after placing an order?
Standard lead time is approximately 3–5 business days when the unit is in stock.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping is available. For delivery to Melbourne, Australia:
• Import duties, customs clearance and GST apply
• Ensure site access: doorways/lifts must accommodate ~60 cm width × ~60 cm depth and allow front clearance of ~114 cm for door swing
• Electrical compatibility: This model is rated at 115 V AC / 60 Hz (US version); for Australia (230 V / 50 Hz) you must order the correct export variant or use a certified step-down transformer approved by an electrician.
How is the unit packaged for transit and what site requirements should be met?
The unit ships boxed and palletized; unpacked footprint is ~81 cm H × ~60 cm W × ~60 cm D, with door-open depth ~114 cm. Ensure a level, firm delivery path from vehicle to installation location and adequate clearance for the door.
What should I do if my shipment arrives damaged or delayed?
On delivery inspect the outer packaging and the unit for visible damage. Photograph any damage, retain packaging and delivery receipt, and contact your supplier with order and tracking details to initiate a freight claim or arrange replacement.
Application FAQ
How do I install and set up the VT65MLMEDADA?
1. Position the unit on a firm, level indoor surface in a climate-controlled room or under-counter space; allow ventilation behind/top and front clearance of ~114 cm for full door opening.
2. Connect to a dedicated grounded outlet matching your region’s voltage.
3. Power on the unit, set the thermostat to the target lower-temperature range (approx. –25 °C) and allow several hours (typically 4–6) for the internal temperature to stabilize before loading sensitive materials.
What best practices should I follow when using this freezer?
• Arrange stored items with spacing on the pull-out drawers to allow airflow; avoid over-packing.
• Label and rotate stored contents (batch number, date/time) where applicable.
• Minimize door-open events to maintain low and uniform temperature.
• Regularly inspect door gasket/seal, hinges/lock mechanism; ensure unit remains level and interior is clean.
What should I do if a temperature excursion or power interruption occurs?
If internal temperature rises above or falls below the supported range or a power interruption occurs:
• Record the event: date/time, observed temperature deviation, and impacted contents.
• Transfer critical items to validated backup freezer storage until conditions stabilize.
• Investigate the root cause (door left open, ambient thermal load increase, power or sensor fault).
• Once internal temperature returns to the target range and stabilizes, verify stored contents for integrity before resuming standard use.
What maintenance and verification routines are recommended?
• Monthly: Inspect door gasket/seal, hinges/lock; ensure unit remains level.
• Quarterly: Clean interior and exterior surfaces; verify thermostat/alarm functioning.
• Periodically: Review any stored temperature-log data, confirm monitoring equipment calibration and retain documented maintenance/service records for compliance.