Description
Summit Compact Laboratory Refrigerator Description:
Summit ARG1ML Compact Laboratory Refrigerator 1.0 cu.ft. with Glass Door is designed and purpose-built for pharmacy, medication, and vaccine applications to support meeting CDC/VFC guidelines that measures 21.5″ H x 17.5″ W x 19.75″ D.
Summit Compact Laboratory Refrigerator Overview:
- Intelligent microprocessor digital temperature controller
- Improved temperature accuracy with a separate temperature probe for continuous product readout to simulate stored product temperature
- Additional sensor reads interior air temperature for superior temperature control
- Adjustable operating control range from +2 °C to +8°C (+35.6 ° F to +46.4° F)
- Digital display of the min/max temperature (updated in 15 minute increments)
- Temperature viewable in Celsius or Fahrenheit to the nearest degree
- Audible/visual alarms for high/low temperature excursions
- Optimized forced air cooling for excellent temperature stability and uniformity with rapid recovery after door openings and product loading
Accucold’s Med-Lab Performance Series all-refrigerators are designed and purpose-built to meet the demanding needs of life science, laboratory, and medical cold storage applications.
The Summit ARG1ML Compact Laboratory Refrigerator is a 1 cu.ft. all-refrigerator with an adjustable control range from +2 to +10ºC. The microprocessor temperature controller is externally located to minimize door openings, helping to protect stored contents from ambient temperature. The buffered temperature probe is encased in a glycol-filled bottle to better simulate temperature of the stored product, with an additional sensor that reads air temperature to ensure superior temperature control. The current and min/max temperature display is viewable in Celsius or Fahrenheit to the nearest tenth of a degree. The audible and visual temperature alarm sounds if the unit’s interior goes out of range. Additional alarms include a power failure alert, door open alarm, and sensor failure alarm. This unit is equipped with remote alarm contacts in the rear and a 1/2″ probe port on the right side to accept additional monitoring devices.
Inside, the Summit ARG1ML Compact Laboratory Refrigerator utilizes adjustable cycle defrost operation to avoid temperature spikes. Optimized forced air cooling with two internal fans ensures excellent temperature stability and uniformity with rapid recovery after door openings. The adjustable plastic-coated wire shelves can be positioned at 1/2″ increments to accommodate virtually any sized item. This unit includes LED internal lighting with an on/off rocker switch. The interior is constructed from white powder coated metal to ensure added durability and better temperature retention.
The Summit ARG1ML Compact Laboratory Refrigerator comes in a white cabinet finish with a blue accented control panel and easy-grip handle. The double pane tempered glass door provides a full view of the interior. The user-reversible door ships with a right hand swing and self-closing function (note: self-closing is disabled once reversed to LHD). This model includes a factory-installed lock and ships with two keys for staff use. Powerful magnetic gaskets ensure a positive seal.
Designed with hydrocarbon refrigerant, the Summit ARG1ML Compact Laboratory Refrigerator provides an energy efficient solution to laboratory and life science storage. This unit is Pyxis, Omnicell, and AcuDose RX compatible. Additional sizes are available. Explore the full Med-Lab series at accucold.com/medlab.
Summit Compact Laboratory Refrigerator Features:
- Purpose-built design– Designed and purpose-built for pharmacy, medication, and vaccine applications to support meeting CDC/VFC guidelines
- +/-1ºC Variation– The published temperature variation is derived from the maximum deviation of an NTC sensor in a 1 oz. vial located nearest the chamber geometric center during a 24 hour test period
- Adjustable temperature– Intuitive controls let you adjust the interior temperature between +2 to +8ºC
- Temperature alarm– Visual and audible alarms alert if the interior temperature rises or falls out of set range
- Remote alarm contacts– Electronic controller is equipped with a volt-free contact for alarm forwarding to an external remote system (contacts located in the external rear of the unit)
- Automatic defrost with adjustable cycle defrost– Adjustable cycle defrost operation ensures the unit avoid temperature spikes and ice buildup
- Access port– Includes a pre-installed 1/2″ probe hole in the right interior wall to allow easier connection to user-provided monitoring equipment
- Open door alarm– Alarm will sound if the door is left ajar for over 4 minutes (adjustable from 1-99 minutes)
- Audible power failure alarm– Battery-powered alarm sounds if the power connection is interrupted
- Sensor failure alarm– Audible and visual alarm warns if the sensor is disrupted
- Password protection– Control panel requires password entry to adjust the parameters beyond the set point
- Microprocessor control panel– Digital controls with a large LED readout in Celsius or Fahrenheit for intuitive temperature management
- Resettable min/max display– Control panel provides a readout of the low and high temperatures in 15 minute increments
- Factory-installed lock– Keyed lock conveniently located towards the top of the door for added security (2 keys included)
- Adjustable plastic coated wire shelves for improved circulation– Multiple shelf positions spaced at 1/2″ to accommodate virtually any size
- Durable interior– Walls are constructed from white powder coated metal for added strength and improved temperature retention
- Interior light– Energy efficient LED lighting with internal on/off switch
- Self-closing door– Added protection in busy settings; NOTE: door is not self-closing once reversed to an LHD swing
- Double pane tempered glass door– Provides a full display of stored contents
- Reversible door– User-reversible door swing for added flexibility
- Easy-grip handle– 17 1/2″ long handle in blue finish for easy accessibility
- Magnetic door gasket– Quality door gaskets ensure a tight seal for energy savings and improved performance
- Extended warranty– 2 years parts and labor/5 years compressor
- Eco-Friendly– Using environmentally friendly hydrocarbons refrigerants for a greener footprint
- Usage stickers– Ships with removable warning stickers for “DO NOT UNPLUG,” “DO NOT ADJUST TEMPERATURE,” and “DO NOT STORE PERSONAL ITEMS”
- Additional choices– Solid and glass door choices available from 1 to 15 cu.ft.
- Versatile design– Compatible with Pyxis®, Omnicell®, and AcuDose RX®
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
Summit Compact Laboratory Refrigerator Overview:
- Height of Cabinet– 21.5″ (55 cm)
- Width– 17.5″ (44 cm)
- Depth with Door Open– 18.0″ (46 cm)
- Depth– 19.75″ (50 cm)
- Depth with Handle– 21.5″ (55 cm)
- Depth with door at 90°– 36.0″ (91 cm)
- Capacity– 1.0 cu.ft. (28 L)
- Defrost Type– Automatic
- Door– Glass
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Amps– 0.6
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 63.0 lbs. (29 kg)
- Shipping Weight– 73.0 lbs. (33 kg)
- Estimated Time to Ship– 1 to 2 days
- Parts & Labor Warranty– 2 Years
- Compressor Warranty– 5 Years
Summit Compact Laboratory Refrigerator Features:
- Door Swing– RHD
- Reversible– Yes
- Shelf Type– Wire
- Shelf Qty– 3
- Adjustable Shelves– Yes
- Thermostat Type– Digital
- Fan Type– Interior
- Refrigerant Type– R600a
- Refrigerant Amount– 0.71oz.
- Interior Light– Yes
- Temperature Range– 2 to 8ºC
Summit Compact Laboratory Refrigerator Dimensions:
- Interior Height– 15.75″ (40 cm)
- Interior Width– 14.0″ (36 cm)
- Interior Depth– 8.0″ (20 cm)
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Medical Professionals
Product FAQ
What is the ARG1ML and what is it used for?
The ARG1ML is a 1 cu ft (~28 L) compact laboratory refrigerator purpose-built for medical, life-science and research-grade cold storage. It supports an interior temperature control range of +2 °C to +10 °C and includes advanced monitoring features.
What are the key technical specifications?
• Capacity: 1.0 cu ft (~28 L).
• External dimensions: Height ~55 cm, Width ~44 cm, Depth ~50 cm; door-open depth ~91 cm.
• Temperature range: +2 °C to +10 °C.
• Shelves: 3 adjustable wire shelves.
• Alarms/monitoring: High/low temperature alarm, door-open alarm, power-failure alarm, sensor-failure alarm, remote alarm contacts; NIST-calibrated data-logger included.
• Warranty: 2 years parts & labor; 5 years compressor.
What safety and security features are included?
Features include:
– Antimicrobial silver-ion coated handle to help reduce germ spread.
– Factory-installed keyed lock for secure access.
– External microprocessor controller and buffered probe to minimize temperature disturbance during door openings.
Which environments is this unit best suited for?
Ideal for laboratories, clinics, pharmacies and research facilities requiring compact, precise refrigeration from +2 °C to +10 °C. Not designed for ultra-low temperature (-20 °C or colder) use unless separately validated.
Shipping FAQ
How long does shipping take after placing an order?
Typical shipping lead time is approximately 5–7 days when the unit is in stock.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping is available. For delivery to Melbourne, Australia you should consider:
• Import duties, customs clearance and GST.
• Site access: confirm doorways/elevators/corridor can accommodate ~44 cm width × ~50 cm depth with at least ~91 cm clearance for door-open.
• Electrical compatibility: This unit is rated at 115 V AC / 60 Hz (U.S. specification) — for Australia (230 V / 50 Hz), ensure you order the correct export variant or use a certified transformer/adaptor.
How is the unit packaged for transit and what site requirements should be met?
Ships boxed and palletized; unpacked footprint: ~55 cm H × 44 cm W × 50 cm D; door-open depth ~91 cm. Delivery area must be level, firm and clear from vehicle to final installation site.
What should I do if my shipment arrives damaged or delayed?
Upon delivery inspect the outer packaging and unit for visible damage. Photograph any damage, retain packaging along with delivery receipt, and contact your supplier with order/tracking details to initiate a freight claim or arrange replacement.
Application FAQ
How do I install and set up the ARG1ML?
1. Place the unit upright on a firm, level indoor surface in a climate-controlled area; allow adequate clearance behind and above for ventilation and approx. 91 cm linear clearance for door-opening.
2. Connect to a dedicated grounded outlet matching the variant ordered (115 V/60 Hz U.S. version or an appropriate local export variant).
3. Power on the unit, set the digital controller between +2 °C and +10 °C, and allow approximately 4–6 hours for the interior temperature to stabilize before loading sensitive contents.
What best practices should I follow when using this refrigerator?
• Place items on the adjustable wire shelves, allowing adequate spacing for airflow; avoid over-packing to maintain uniform cooling.
• Label stored materials (batch number, date/time) and apply first-in/first-out (FIFO) rotation where applicable.
• Minimize door-open time; use the external controller/display and included data-logger rather than frequent door access.
• Regularly inspect door gasket/seal, lock/handle mechanism, and confirm the interior remains clean and the unit remains level.
What should I do if a temperature excursion or power interruption occurs?
If the internal temperature rises above or falls below the target +2 °C to +10 °C range or a power interruption occurs:
• Record the event: date/time, observed temperature, and impacted contents.
• Relocate critical contents to validated backup storage until conditions stabilize.
• Investigate root cause (for example: door left open, ambient heat load, power or sensor fault).
• Once the unit returns to stable operation within the target range, inspect stored items for integrity before resuming standard use.
What maintenance and verification routines are recommended?
• Monthly: Inspect door gasket/seal, hinges and lock/handle operation; ensure the unit remains level.
• Quarterly: Clean interior/exterior surfaces; verify ventilation/exhaust pathways and alarm/data-logger function.
• Periodically: Export and review internal min/max temperature logs, confirm calibration certificate validity, and keep service/maintenance documentation for compliance.