Description
Summit Undercounter ADA Laboratory Refrigerator Description:
Accucold’s Med-Lab Performance Series all-refrigerators are designed and purpose-built to meet the demanding needs of life science, laboratory, and medical cold storage applications, with a powder-coated silver ion handle that helps prevent germ spread.
The Summit ARS6MLDR Undercounter ADA Laboratory Refrigerator with Removable Drawers is a 6 cu.ft. all-refrigerator sized at 32.5 inches high for use in most ADA compliant applications. The adjustable control range allows you to manage the temperature between +2 to +10ºC. The microprocessor temperature controller is externally located to minimize door openings, helping to protect stored contents from ambient temperature. The buffered temperature probe is encased in a glycol-filled bottle to better simulate temperature of the stored product, with an additional sensor that reads air temperature to ensure superior temperature control. The current and min/max temperature display is viewable in Celsius or Fahrenheit to the nearest tenth of a degree. The audible and visual temperature alarm sounds if the unit’s interior goes out of range. Additional alarms include a power failure alert, door open alarm, and sensor failure alarm. This unit is equipped with remote alarm contacts in the rear and a 1/2″ probe port on the right side to accept additional monitoring devices. A hospital grade cord with ‘green dot’ plug is also included for added safety in high traffic areas.
Inside, the Summit ARS6MLDR Undercounter ADA Laboratory Refrigerator utilizes adjustable cycle defrost operation to avoid temperature spikes. Optimized forced air cooling with one internal fan ensures excellent temperature stability and uniformity with rapid recovery after door openings. Three white powder-coated aluminum drawers provide ample storage and easy organization. Positioned on ball-bearing glides, the drawers offer full-extension for complete product access. Additionally, each drawer is removable allowing for transport of contents as needed. The drawers also lock into place for added security, while the perforated bottom ensures proper ventilation throughout the unit. LED internal lighting with an on/off rocker switch attractively illuminates the interior for easy viewing. The interior is constructed from white powder-coated metal to ensure added durability and better temperature retention.
The Summit ARS6MLDR Undercounter ADA Laboratory Refrigerator comes in a white cabinet finish with a blue accented control panel. The factory-installed antimicrobial handle is powder-coated with naturally occurring silver ions, which help to reduce the spread of germs, allowing for a safer, cleaner user experience. The user-reversible door ships with a right-hand swing and self-closing function (note: self-closing is disabled once reversed to LHD). This model includes a factory-installed lock and ships with two keys for staff use. Powerful magnetic gaskets ensure a positive seal.
Designed with hydrocarbon refrigerant, the ARS6MLDR provides an energy-efficient solution to laboratory and life science storage. This unit is Pyxis, Omnicell, and AcuDose RX compatible. Additional sizes, door options, and units with a combination of standard shelving and drawers are also available. Explore the full Med-Lab series at accucold.com/medlab.
NOTE: Accucold refrigerators are purpose-built to the support meeting the current CDC (Centers for Disease Control) guidelines for the storage of vaccines. At this time, Accucold refrigerators have not been evaluated to the new NSF 456 standard. As referenced in the current Vaccine Storage and Handling Toolkit, these units incorporate microprocessor controls and forced air circulation for stable operation. For additional temperature assurance, the CDC recommends a calibrated digital data logger (DDL), which can be added with the factory-installed option DL2B or addition of the DL2BKit accessory.
Summit Undercounter ADA Laboratory Refrigerator Features:
- Purpose-built design– Designed and purpose-built to meet the demanding needs of scientific, life science, laboratory, and medical cold storage applications
- +/-1ºC Variation– The published temperature variation is derived from the maximum deviation of an NTC sensor in a 1 oz. vial located nearest the chamber geometric center during a 24 hour test period
- Antimicrobial silver ion-coated handle– Powder-coated handle with naturally occurring silver ions to help reduce the spread of germs
- Adjustable temperature– Intuitive controls let you adjust the interior temperature between +2 to +10ºC
- Full-extension ventilated drawers– Three drawers made from powder-coated aluminum glide out smoothly for convenient storage, with perforated bottoms to ensure proper air circulation
- Removable drawers– Drawers can be easily removed for convenient transporting of contents to a lab counter or work station
- Temperature alarm– Visual and audible alarms alert if the interior temperature rises or falls out of set range
- Remote alarm contacts– Electronic controller is equipped with a volt-free contact for alarm forwarding to an external remote system (contacts located in the external rear of the unit)
- Automatic defrost with adjustable cycle defrost– Adjustable cycle defrost operation ensures the unit avoid temperature spikes and ice buildup
- Access port– Includes a pre-installed 1/2″ probe hole in the right interior wall to allow easier connection to user-provided monitoring equipment
- Open door alarm– Alarm will sound if the door is left ajar for over 4 minutes (adjustable from 1-99 minutes)
- Audible power failure alarm– Battery-powered alarm sounds if the power connection is interrupted
- Sensor failure alarm– Audible and visual alarm warns if the sensor is disrupted
- Password protection– Control panel requires password entry to adjust the parameters beyond the set point
- Microprocessor control panel– Digital controls with a large LED readout in Celsius or Fahrenheit for intuitive temperature management
- Resettable min/max display– Control panel provides a readout of the low and high temperatures in 15 minute increments
- Factory-installed lock– Keyed lock conveniently located towards the top of the door for added security (2 keys included)
- Durable interior– Walls are constructed from white powder coated metal for added strength and improved temperature retention
- Hospital grade cord with ‘green dot’ plug– Secure three-pronged cord for added unit safety
- Interior light– Energy efficient LED lighting with internal on/off switch
- Self-closing door– Added protection in busy settings; NOTE: door is not self-closing once reversed to an LHD swing
- Reversible door– User-reversible door swing for added flexibility
- Magnetic door gasket– Quality door gaskets ensure a tight seal for energy savings and improved performance
- Extended warranty– 2 years parts and labor/5 years compressor
- Eco-Friendly– Using environmentally friendly hydrocarbons refrigerants for a greener footprint
- Usage stickers– Ships with removable warning stickers for “DO NOT UNPLUG,” “DO NOT ADJUST TEMPERATURE,” and “DO NOT STORE PERSONAL ITEMS”
- Additional choices– Solid and glass door choices available from 1 to 15 cu.ft. Browse the Med-Lab series for more information
- Versatile design– Compatible with Pyxis®, Omnicell®, and AcuDose RX®
- ADA compatible height– Sized at 32.5 inches high for use with most ADA compliant applications
- 100% CFC free– Environmentally friendly design without ozone-damaging chemicals
- CARB Compliant– Environmentally friendly hydrocarbon-based refrigerant for a greener footprint that meets current CARB regulations
Summit Undercounter ADA Laboratory Refrigerator Overview:
- Height of Cabinet– 32.5″
- Width– 23.38″
- Width with Door Open– 23.88″
- Depth– 24.63″
- Depth with Handle– 26.13″
- Depth with door at 90°– 46.5″
- Capacity– 6.0 cu.ft.
- Defrost Type– Automatic
- Door– White
- Cabinet– White
- US Electrical Safety– ETL
- Canadian Electrical Safety– ETL-C
- Amps– 0.75
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 130.0 lbs.
- Shipping Weight– 141.0 lbs.
- Parts & Labor Warranty– 2 Years
- Compressor Warranty– 5 Years
Summit Undercounter ADA Laboratory Refrigerator Features:
- Door Swing– RHD
- Reversible– Yes
- Shelf Type– Drawers
- Shelf Qty– 3
- Adjustable Shelves– Yes
- Thermostat Type– Digital
- Fan Type– Interior
- Refrigerant Type– R600a
- Refrigerant Amount– 1.66oz.
- Interior Light– Yes
- Temperature Range– 2 to 10ºC
- Sound Level– 36.4 dB
Summit Undercounter ADA Laboratory Refrigerator Dimensions:
- Interior Height– 24.5″
- Interior Width– 20.0″
- Interior Depth– 17.0″
- Compressor Step Height– 6.5″
- Compressor Step Width– 20.0″
- Compressor Step Depth– 5.0″
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the ARS6MLDR and what is it designed for?
The ARS6MLDR is an ADA-height undercounter laboratory refrigerator with approximately 170 L (≈ 6 cu ft) of storage capacity and exterior dimensions of roughly 83 cm H × 59 cm W × 63 cm D (≈ 32.5″ × 23.4″ × 24.8″). It is engineered for medical, pharmaceutical, and laboratory environments that require compact, precise cold storage within ADA height compliance.
What temperature range and control features does it support?
The internal temperature range is approximately +2 °C to +10 °C (≈ 36 °F to 50 °F).
It includes a digital microprocessor thermostat with an external temperature display, a glycol-filled buffered probe to simulate stored-product conditions, and a secondary air sensor for increased temperature stability and accuracy.
What alarms, monitoring and security features are included?
The ARS6MLDR provides visual and audible alarms for high/low temperature, door ajar, power failure, and sensor error.
Additional features include a remote alarm contact output, a 12.7 mm (½″) probe access port for external monitoring devices, and a keyed lock for secure content protection.
What construction and usability features support its laboratory application?
The unit features a self-closing reversible door with an antimicrobial silver-ion handle, full-extension ventilated drawers on ball-bearing glides for easy organization, LED interior lighting, and forced-air cooling for uniform temperature.
It has a sealed-back powder-coated steel cabinet designed for easy cleaning and long-term reliability in laboratory environments.
Shipping FAQ
What are the exterior dimensions and shipping weight?
Approximate size: 83 cm H × 59 cm W × 63 cm D.
Typical shipping weight is around 64 kg (≈ 140 lb).
What electrical and installation requirements apply?
Operates on 115 V AC / 60 Hz and draws roughly 1.4 amps.
Requires a dedicated grounded outlet, a level surface, and sufficient clearance for door swing and ventilation — particularly when installed under counters or built into cabinetry.
What steps should be taken during unpacking and placement?
Inspect for visible transit damage upon delivery. If shipped or stored on its side, allow several hours upright before powering on.
Ensure the unit is level, confirm adequate airflow for ventilation, and verify that the controller and alarms are functioning before adding contents.
What delivery or shipping method is typical for this model?
This professional-grade undercounter unit typically ships via palletized freight.
Prepare a level installation site, ensure clear access for delivery, and confirm the installation area meets height and ventilation requirements.
Application FAQ
Is this unit suitable for pharmaceutical or clinical cold-storage applications?
Yes — the ARS6MLDR is ideal for use in laboratories, clinics, and healthcare facilities. It supports the +2 °C to +10 °C range and includes the monitoring, alarm, and temperature-control features needed for regulated cold storage.
What daily best-practice steps should be followed for operation?
Use a calibrated data logger at mid-drawer height, label and rotate stored samples, avoid over-packing to maintain airflow, minimize door-open time, and document daily minimum/maximum temperature readings and alarms.
What routine maintenance is recommended?
Clean condenser coils and ventilation openings twice yearly, test alarms and temperature controls monthly, check gaskets and drawer seals annually, and keep the unit level with unrestricted airflow.
What should be done in the event of a temperature excursion or alarm?
Record the date/time, the temperature deviation, and any affected materials.
Move critical samples to backup storage if required, correct the issue (e.g., door left open, blocked vent, power loss), and restore conditions to within the desired range before restocking.