Description
Summit -85ºC Ultra Low Chest Freezer Description:
Accucold introduces a new series of ultra low temperature chest freezers designed for reliable storage at -85ºC for laboratory, comes with 9 Vertical box shelves.
Manufactured for Accucold in Denmark by ELCOLD, model LAB11R9 offers 4.8 cu.ft. of storage. Extra thick 4 1/2″ insulation and static manual defrost operation help to maintain ultra low temperatures suitable for maintaining the integrity of biological samples, industrial materials, laboratory specimens, select vaccines, and other materials with highly specific cold storage requirements. Two inner lids offer added protection by shielding contents when the outer lid is opened. The digital microprocessor controls are mounted in the front for convenient access. This unit has a temperature range that can be set between -50 and -85ºC, with a display viewable in Celsius or Fahrenheit. There is also an audible 9V battery-powered alarm in the rear of the unit that will sound in the event of power failure.
For easy mobility, the Summit LAB11R9 4.8 Cu.Ft. -85ºC Ultra Low Chest Freezer comes with pre-installed casters, including two front locking wheels to help keep the unit in place. The easy-grip handle includes a factory-installed keyed lock for added security. We include a set of 9 vertical stainless steel box racks (6 10-box and 3 5-box) for additional storage convenience.
The LAB11 is UL listed for use in the US and Canada, and also meets ANSI/NSF-7. Additional sizes will be available soon.
Summit -85ºC Ultra Low Chest Freezer Features:
- Ultra Low Temperature -85ºC Chest Freezer– Designed for highly specialized cold storage storage applications of biological samples, DNA/RNA material, chemicals, select vaccines and antibiotics, and other substances with strict low temperature requirements
- Advanced design– Interior walls feature more than 8 layers of D-shaped cooling pipes to ensure increased contact with the freezer walls in order to ensure maximum insulation and product efficiency
- Manufactured by Elcold for Accucold– Designed and constructed in Denmark
- Inner lids– Includes two removable inner lids for added protection from ambient air when opening the outer lid
- Extra thick insulation– 4 1/2″ thick walls for added efficiency and stable cooling
- 9 Vertical box shelves included– Ships with 6 10-box shelves and 3 5-box shelves in stainless steel for added storage convenience
- Factory installed lock– Exterior lock on the handle for reliable security (2 keys included)
- Digital thermostat– Electronic microprocessor controls for precise temperature management, with temperature readout and controls externally located for added convenience
- ºC/ºF Readout– Temperature display can be alternated between Celsius or Fahrenheit to best suit your preference
- High/low temperature alarm– Audible alarm warns if the interior temperature goes out of the set range
- Counterbalanced hinges– Easy and safe loading and unloading with a lid that remains open when lifted past 45º angles
- Pre-installed casters– Four wheels (front two locking) to allow easier mobility when cleaning or moving the unit
- Manual defrost– Static cooling for a stable interior
- Defrost drain– Front-located drain helps to make the defrosting process more convenient
- Steel interior– Durable construction in a white finish for easy visibility
- Power-out alarm– 9V battery-powered audible alarm mounted in the rear sounds in the event of power failure
- Hermetically sealed compressor– Ensures lasting durability and reduced maintenance in busy facilities
- Approved for use in USA and Canada– UL and UL-C listed
- Approved to ANSI/NSF-7– UL-S listed to the National Sanitation Foundation standard
- 100% CFC Free– Environmentally friendly design with no ozone-damaging chemicals
- CARB Compliant– Environmentally friendly proprietary blend of hydrocarbon-based refrigerant for a greener footprint that meets current CARB regulations
Summit -85ºC Ultra Low Chest Freezer Specifications:
- Height of Cabinet– 36.25″
- Width– 28.5″
- Depth– 28.0″
- Depth with Handle– 30.0″
- Capacity– 4.8 cu.ft.
- Defrost Type– Manual
- Door– White
- Cabinet– White
- US Electrical Safety– UL
- Canadian Electrical Safety– ULC
- Sanitation– UL-S
- Amps– 5.3
- Voltage/Frequency– 115 V AC/60 Hz
- Weight– 150.0 lbs.
- Shipping Weight– 160.0 lbs.
- Parts & Labor Warranty– 1 Year
- Compressor Warranty– 5 Years
Summit -85ºC Ultra Low Chest Freezer Dimensions:
- Door Swing– Lift-Up
- Interior Height– 26.25″
- Interior Width– 20.25″
- Interior Depth– 17.5″
- Thermostat Type– Digital
- High Side PSI– 400.0
- Low Side PSI– 174.0
- Wheels– 4
- Compressor Step Height– 9.75″
- Compressor Step Width– 7.75″
- Compressor Step Depth– 17.5″
- Temperature Range– -50 to -85ºC
- Sound Level– 49 dB
Summit Appliance manufactures compact and specialty major appliances: including refrigerators, freezers, laboratory and scientific refrigeration and refrigerated and freezer drawers. Our products are grouped into: laboratory, medical and scientific refrigerators and freezers.
Product FAQ
What is the LAB11R9 and what is it used for?
The LAB11R9 is an ultra-low temperature chest freezer (4.8 cu ft / ~136 L) capable of reaching as low as approximately -85 °C, designed for laboratory, life-science, and medical storage applications including biological samples, chemicals and select vaccines/antibiotics.
What are the key technical specifications?
• Capacity: 4.8 cu ft (~136 L).
• External dimensions: Height ~92 cm (36.25″), Width ~72 cm (28.5″), Depth ~71 cm (28.0″); depth with handle ~76 cm (30.0″).
• Temperature range: approximately -50 °C to -85 °C (preset around -20 °C to -85 °C).
• Defrost type: Manual.
• Electrical: 115 V AC / 60 Hz; Amps around 5.3. Weight ~68 kg (~150 lbs).
What safety and monitoring features are included?
Features include extra thick insulation (~4½″ thick walls) for stable ultra-low temperatures, a digital microprocessor thermostat with external display, high/low temperature alarm, power-out (battery-backed) alarm, factory lock on the handle, pre-installed casters (two front locking) for mobility, and dual inner lids to minimize warm air intrusion when loading.
Which environments is this unit best suited for?
Ideal for hospital labs, research facilities, biobanks or other healthcare settings requiring ultra-low temperature storage (-50 °C to -85 °C) for biological samples or specialized materials. Not appropriate for standard domestic household freezing needs.
What warranty coverage is provided?
Manufacturer’s warranty: 1 year parts & labor; 5 years compressor.
Shipping FAQ
How long does shipping take after placing an order?
The typical shipping lead time is 1-2 days for dispatch when in stock (U.S. region) as listed for this model.
Is international shipping available and what should I consider for Melbourne, Australia?
Yes — international shipping can be arranged. For delivery to Melbourne, Australia you should consider:
• Import duties, customs clearance & Goods & Services Tax (GST).
• Freight access: ensure door width, corridor/elevator clearance and appropriate floor loading capacity.
• Electrical/mains compatibility: This model is specified for **115 V AC/60 Hz** (U.S. version) — if installing in Australia (230 V /50 Hz), ensure you order the correct variant or provide a suitable step-down transformer.
How is the unit packaged for transit and what site requirements should be met?
The unit ships palletized with protective packaging. Approx. unpacked footprint: height ~92 cm × width ~72 cm × depth ~71 cm; allow additional clearance for lid/lift operation. Delivery site should have a firm level surface, adequate clearance for entry and for opening the lid, and handling equipment (pallet jack/lift) if required.
What should I do if my shipment arrives damaged or is delayed?
On delivery, inspect the packaging and unit for visible damage or transit delay. If damage is present: photograph the condition, retain the packaging and delivery receipt, and contact your supplier with your order/tracking details to initiate a claim or replacement.
Application FAQ
How do I install and set up the LAB11R9?
1. Position the freezer on a firm, level surface in a climate-controlled room. Allow clearance behind and above as specified by the installer for ventilation and to ensure lid-opening clearance.
2. Connect to a dedicated grounded outlet matching the unit’s specification (115 V AC/60 Hz). For Australian installations ensure correct variant or transformer.
3. Power on the unit, set the target temperature (within the designed range e.g., -50 °C to -85 °C). Allow several hours (typically 3-6 h) for the internal temperature to stabilize before loading sensitive materials.
4. Verify that the control panel/display is working, alarms (high/low temperature, power‐out) are operational, the lock and handle mechanism operate smoothly, and the casters/level are correct prior to routine use.
What best practices should I follow when storing biological samples or ultra-cold materials?
• Use the included stainless steel vertical box racks to organize samples (6×10-box shelves + 3×5-box shelves).
• Minimize lid openings and duration of exposure to ambient air; keep frequently accessed items near the front.
• Clearly label stored items with date/time, sample ID or batch number to support traceability.
• Regularly inspect the lid gasket/seal, hinges and lock mechanism; maintain logs of internal minimum/maximum temperature and alarm events if required by your facility protocol.
What should I do if a temperature excursion or power interruption occurs?
If the internal temperature rises above or falls below the required safe range, or if a power interruption occurs:
• Record the event including date/time, observed temperature and impacted contents.
• Immediately relocate critical contents to validated backup ultra-low temperature storage until conditions stabilize.
• Investigate underlying cause (lid left open, ambient heat load, sensor/power fault).
• Once the freezer returns to normal operation and internal temperature is stable within range, verify sample integrity and conditions before returning materials.
What maintenance and verification routines are recommended?
• Monthly: Inspect lid gasket/seal, hinges, locking handle and caster/level condition.
• Quarterly: Clean condenser coils (if externally accessible) and ventilation pathways; ensure airflow and fan systems are unobstructed.
• Periodically: Test audible/visual alarms, power-out battery alarm, any remote monitoring ports; perform sensor calibration if required. Keep service, calibration and alarm logs for audit readiness.